Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s
TRIGger:WINDow:MINWidth <MinDuration>
Sets the lower time limit of the stay inside or outside the window (comparisons WITHin
and OUTSide).
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:WINDow:MAXWidth <MaxDuration>
Sets the upper time limit of the stay inside or outside the window (comparisons WITHin
and OUTSide).
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.4 Waveform Analysis
15.4.1 Zoom
See also Chapter 4.1, "Zoom", on page 78.
ZOOM:ENABle.............................................................................................................. 338
ZOOM:SCALe............................................................................................................... 339
ZOOM:POSition.............................................................................................................339
ZOOM:ENABle <Enabled>
Enables or disables the zoom.
Parameters:
<Enabled> ON | OFF
*RST: OFF
Waveform Analysis