Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
334User Manual 1326.1578.02 ─ 15
Parameters:
<Pattern> String that contains the pattern in binary format. The parameter
accepts the bit value X (don't care).
Example:
TRIGger:SPATtern:PATTern '11001100'
TRIGger:SPATtern:PATTern '110011XX'
15.3.5.13 Timeout Trigger (R&S RTH-K19)
TRIGger:TIMeout:RANGe...............................................................................................334
TRIGger:TIMeout:TIME.................................................................................................. 334
TRIGger:TIMeout:RANGe <TimeoutMode>
Selects the relation of the signal level to the trigger level, which is specified with
TRIGger:LEVel<m>:VALue
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
*RST: HIGH
TRIGger:TIMeout:TIME <Time>
Sets the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.14 Interval Trigger (R&S RTH-K19)
TRIGger:INTerval:SLOPe................................................................................................334
TRIGger:INTerval:RANGe............................................................................................... 335
TRIGger:INTerval:WIDTh................................................................................................ 335
TRIGger:INTerval:DELTa.................................................................................................335
TRIGger:INTerval:MINWidth............................................................................................ 335
TRIGger:INTerval:MAXWidth...........................................................................................336
TRIGger:INTerval:SLOPe <Slope>
Sets the edge direction for the trigger. You can analyze the inteval between positive
edges or between negative edges.
Parameters:
<Slope> POSitive | NEGative
*RST: POSitive