Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
331User Manual 1326.1578.02 ─ 15
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
See also Chapter 3.6.11, "Data2Clock Trigger (R&S RTH-K19)", on page 69.
TRIGger:DATatoclock:DSOurce[:VALue]...........................................................................331
TRIGger:DATatoclock:CSOurce[:VALue]...........................................................................331
TRIGger:DATatoclock:CSOurce:EDGE.............................................................................331
TRIGger:DATatoclock:CONDition.....................................................................................331
TRIGger:DATatoclock:HTIMe...........................................................................................332
TRIGger:DATatoclock:STIMe...........................................................................................332
TRIGger:DATatoclock:DSOurce[:VALue] <Source>
Selects the input channel of the data signal.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:DATatoclock:CSOurce[:VALue] <Source>
Selects the input channel of the clock signal.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:DATatoclock:CSOurce:EDGE <Slope>
Sets the edge of the clock signal: rising (POSitive), falling (NEGative), or both
edges (EITHer). The time reference point for the setup and hold time is the crossing
point of the clock edge and the trigger level.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:DATatoclock:CONDition <TrigCondition>
Selects how a violation of the setup and hold time is handled.
Parameters:
<TrigCondition> VIOLation | OK
Waveform Setup