Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
329User Manual 1326.1578.02 ─ 15
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
See also Chapter 3.6.10, "Slew Rate Trigger (R&S RTH-K19)", on page 68.
TRIGger:LEVel<m>:SLEW:LOWer...................................................................................329
TRIGger:LEVel<m>:SLEW:UPPer....................................................................................329
TRIGger:SLEW:SLOPe.................................................................................................. 329
TRIGger:SLEW:RANGe..................................................................................................329
TRIGger:SLEW:RATE.....................................................................................................330
TRIGger:SLEW:DELTa................................................................................................... 330
TRIGger:SLEW:MINWidth...............................................................................................330
TRIGger:SLEW:MAXWidth..............................................................................................330
TRIGger:LEVel<m>:SLEW:LOWer <Level>
TRIGger:LEVel<m>:SLEW:UPPer <Level>
Set the upper and lower voltage thresholds, respectively. The time measurement starts
when the signal crosses the first trigger level - the upper or lower level depending on
the selected slope - and stops when the signal crosses the second level.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger:SLEW:SLOPe <Slope>
Sets the edge direction for the trigger.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:SLEW:RANGe <Condition>
Defines the time limits of the slew rate.
Parameters:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
Waveform Setup