Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
327User Manual 1326.1578.02 ─ 15
TRIGger:LEVel<m>:RUNT:LOWer................................................................................... 327
TRIGger:LEVel<m>:RUNT:UPPer.................................................................................... 327
TRIGger:RUNT:POLarity.................................................................................................327
TRIGger:RUNT:RANGe.................................................................................................. 327
TRIGger:RUNT:WIDTh................................................................................................... 328
TRIGger:RUNT:DELTa....................................................................................................328
TRIGger:RUNT:MINWidth............................................................................................... 328
TRIGger:RUNT:MAXWidth.............................................................................................. 328
TRIGger:LEVel<m>:RUNT:LOWer <Level>
TRIGger:LEVel<m>:RUNT:UPPer <Level>
Set the upper and lower levels that limit the runt.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger:RUNT:POLarity <Polarity>
Sets the pulse polarity, that is the direction of the first pulse slope. You can trigger on:
Positive going pulses. The width is defined from the rising to the falling edge.
Negative going pulses. The width is defined from the falling to the rising edge.
Both positive and negative going pulses
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:RUNT:RANGe <Condition>
Defines an additional time limit of the runt pulse.
Parameters:
<Condition> ANY | LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer | SHORter
Defines a minimum time during which the signals match the pat-
tern definition. The minimum time is defined by
Waveform Setup