Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
325User Manual 1326.1578.02 ─ 15
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:PATTern:WIDTh:DELTa <Tolerance>
Sets a range Δt to the pattern width that is defined by TRIGger:PATTern:WIDTh[:
WIDTh].
Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s
TRIGger:PATTern:WIDTh:MINWidth <MinDuration>
Sets the lower time limit of the pattern match for range conditions WITHin and OUT-
Side.
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:PATTern:WIDTh:MAXWidth <MaxDuration>
Sets the upper time limit of the pattern match for range conditions WITHin and OUT-
Side.
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.8 State Trigger (R&S RTH-K19)
See also Chapter 3.6.8, "State Trigger (R&S RTH-K19)", on page 65.
TRIGger:STATe:CHANnel<m>.........................................................................................326
TRIGger:STATe:COMBination..........................................................................................326
TRIGger:STATe:CSOurce[:VALue]................................................................................... 326
TRIGger:STATe:CSOurce:EDGE..................................................................................... 326
Waveform Setup