Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
323User Manual 1326.1578.02 ─ 15
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
15.3.5.7 Pattern Trigger (R&S RTH-K19)
See also Chapter 3.6.7, "Pattern Trigger (R&S RTH-K19)", on page 63.
TRIGger:PATTern:STATe[:CHANnel<m>].......................................................................... 323
TRIGger:PATTern:STATe:COMBination.............................................................................323
TRIGger:PATTern:WIDTh:RANGe....................................................................................324
TRIGger:PATTern:TIMeout[:TIME]....................................................................................324
TRIGger:PATTern:WIDTh[:WIDTh]................................................................................... 324
TRIGger:PATTern:WIDTh:DELTa..................................................................................... 325
TRIGger:PATTern:WIDTh:MINWidth.................................................................................325
TRIGger:PATTern:WIDTh:MAXWidth................................................................................325
TRIGger:PATTern:STATe[:CHANnel<m>] <State>
Sets the state of each input channel. The channel is specified by the channel suffix:
The logical combination of the channel states is defined by TRIGger:PATTern:
STATe:COMBination.
Suffix:
<m>
.
1..4: analog channel 1 to 4
8..15: digital channels D0 to D7
Parameters:
<State> ONE | ZERO | DONTcare
ONE
The signal value is above the defined threshold.
ZERO
The signal value is below the defined threshold.
DONTcare
The signal state does not matter.
*RST: DONTcare
TRIGger:PATTern:STATe:COMBination <ChCombination>
Sets the logical combination for all active channels. The required state of each channel
is defined by TRIGger:STATe:CHANnel<m>.
Parameters:
<ChCombination> AND | OR
*RST: AND