Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
316User Manual 1326.1578.02 ─ 15
Parameters:
<Time> Range: 8E-9 to 10
Increment: 200E-6
*RST: 1E-3
Default unit: s
TRIGger:HOLDoff:EVENts <Events>
Sets the number of triggers to be skipped until the next trigger occurs. The command
takes effect if TRIGger:MODE is set to EVENts.
Parameters:
<Events> Range: 1 to 1000000000
Increment: 10
*RST: 1
TRIGger:HOLDoff:MIN
<RandomMinTime>
TRIGger:HOLDoff:MAX <RandomMaxTime>
Set the time limits for random holdoff time. For each acquisition, the instrument selects
a new random holdoff time from the specified range.
Parameters:
<RandomMinTime>
<RandomMaxTime>
Range: 8E-9 to 10
Increment: 200E-6
*RST: 2E-3
Default unit: s
TRIGger:MNR <MoreNoiseReject>
Enables a hysteresis to avoid unwanted trigger events caused by noise oscillation
around the trigger level.
Parameters:
<MoreNoiseReject> ON | OFF
*RST: OFF
15.3.5.2 Edge Trigger
TRIGger:EDGE:SLOPe <Slope>
Sets the edge to be triggered on.
Parameters:
<Slope> POSitive | NEGative | EITHer
POSitive = rising edge
NEGative = falling edge
EITHer = rising and falling edge
*RST: POSitive
Waveform Setup