Instructions

Table Of Contents
Remote Control Commands
R&S
®
Scope Rider RTH
314User Manual 1326.1578.02 ─ 15
TRIGger:HOLDoff:MIN....................................................................................................316
TRIGger:HOLDoff:MAX.................................................................................................. 316
TRIGger:MNR................................................................................................................316
TRIGger:MODE <Mode>
The trigger mode determines the behavior of the instrument if no trigger occurs, and
also the number of acquired waveforms when a trigger occurs.
Parameters:
<Mode> AUTO | NORMal | SINGle
AUTO
The instrument triggers repeatedly after a time interval if the trig-
ger conditions are not fulfilled. If a real trigger occurs, it takes
precedence.
NORMal
The instrument acquires waveforms continuously, each time
when a trigger occurs.
SINGle
When a trigger occurs, the instrument acquires one waveform.
*RST: AUTO
TRIGger:SOURce <Source>
Selects the trigger source, the waveform on which the trigger condition is checked.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
Logic channels D0..D7 require option R&S RTH-B1. For video,
runt, slew rate and window trigger, only analog channels are
available.
*RST: C1
TRIGger:TYPE <Type>
Selects the trigger type, the event type that defines the trigger point.
Parameters:
<Type> EDGE | GLITch | WIDTh | TV | PATTern | STATe | RUNT |
SLEWrate | DATatoclock | SERPattern | TIMeout | INTerval |
WINDow | PROTocol
EDGE | GLITch | WIDTh | TV
Standard trigger types
PATTern | STATe | RUNT | SLEWrate | DATatoclock | SERPat-
tern | TIMeout | INTerval | WINDow
Require option R&S RTH-K19
PROTocol
Requires option R&S RTH-K1 and/or R&S RTH-K2
Waveform Setup