Remote Control Commands
R&S
®
Scope Rider RTH
312User Manual 1326.1578.02 ─ 15
MIDDle
Limits the record length to 12.5 kSample.
MIN
Limits the record length to 1.25 kSample.
*RST: MAX
The actual record length can differ from these limits, see "Act. Record Len. C1 - C4"
on page 49.
ACQuire:WAVeformupd <WaveformUpdate>
The command is relevant if the time scale is ≥50 ms/div. At these slow timebases, you
can select how the acquired samples are displayed.
Parameters:
<WaveformUpdate> INTermediate | FULL
INTermediate
The acquired samples are displayed before the acquisition is
completed. In this mode, the record length is limited to
125 kSamples.
FULL
The acquired samples are displayed when the complete acquisi-
tion has been recorded. Depending on the selected time scale, it
takes some time until the waveform is visible. This mode does
not limit the record length and is always used for time scales
<50 ms/div.
*RST:
INTermediate
ACQuire:POINts:ARATe?
Returns the number of captured analog waveform points per second (sample rate of
the ADC).
Return values:
<ADCSampleRate> 1 active channel: 5 Gsample/s
2 active channels: 2.5 Gsample/s
3 or 4 active channels: 1.25 Gsample/s
Default unit: Sa/s
Usage: Query only
ACQuire:POINts[:VALue]?
Returns the record length, number of captured waveform samples before interpolation.