309User Manual 1326.1578.02 ─ 15
Return values:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
CHANnel<m>:THReshold:FINDlevel
The instrument sets the threshold for the selected channel.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Usage: Event
15.3.3 Horizontal Setup
TIMebase:SCALe...........................................................................................................309
TIMebase:RANGe..........................................................................................................309
TIMebase:HORizontal:POSition.......................................................................................309
TIMebase:REFerence.....................................................................................................310
TIMebase:SCALe <Scale>
Sets the time scale of the horizontal axis for all signals.
Parameters:
<Scale> Range: 1E-9 to 500
Increment: Steps 1, 2, 5 (1, 2, 5, 10, 20, 50, 100, 200, 500...)
*RST: 100E-9
Default unit: s/div
TIMebase:RANGe <AcquisitionTime>
Sets the acquisition time, the timerange across the 10 horizontal divisions of the dia-
gram. Use the command alternatively to TIMebase:SCALe.
Parameters:
<AcquisitionTime> Range: 10E-9 to 5000
Increment: Steps 1, 2, 5 (1, 2, 5, 10, 20, 50, 100, 200, 500...)
*RST: 1E-6
Default unit: s
TIMebase:HORizontal:POSition <Position>
Sets the horizontal position of the trigger point in relation to the reference point.
See also: "Horizontal Position" on page 46