CHANnel<m>:THReshold:TECHnology <ThresholdType>
Sets the threshold value for digitization of analog signals. If the signal value is higher
than the threshold, the signal state is high (1 or true for the boolean logic). Otherwise,
the signal state is considered low (0 or false) if the signal value is below the threshold.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
GND
0 V (for CAN channels, requires option R&S RTH-K3)
CAN
2 V (for CAN channels, requires option R&S RTH-K3)
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V (for LIN channels, requires option R&S RTH-
K3)
USER
Set the value with CHANnel<m>:THReshold:USER.
*RST:
TTL
CHANnel<m>:THReshold:USER <ThresholdValue>
Set an individual threshold value if CHANnel<m>:THReshold:TECHnology is set to
USER.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
CHANnel<m>:THReshold:THReshold?
Returns the threshold value.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Waveform Setup