307User Manual 1326.1578.02 ─ 15
CHANnel<m>:BANDwidth <BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<BandwidthLimit> FULL | B350 | B200 | B100 | B60 | B50 | B40 | B20 | B10 | B5 |
B4 | B2 | B1 | B5HK | B4HK | B2HK | B1HK | B50K | B40K |
B20K | B10K | B5K | B4K | B2K | B1K
FULL
At full bandwidth, all frequencies in the specified range are
acquired and displayed.
B350 | B200 | B100 | B60 | B50 | B40 | B20 | B10 | B5 | B4 | B2
| B1
Limit to 350 MHz, 200 MHz, 100 MHz, 60 MHz,..., respectively.
B5HK | B4HK | B2HK | B1HK | B50K | B40K | B20K | B10K |
B5K | B4K | B2K | B1K
Limit to 500 kHz, 400 kHz,..., respectively.
*RST:
FULL
CHANnel<m>:POLarity <Polarity>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the horizontal display center. Inversion
affects only the display of the signal but not the trigger.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
CHANnel<m>:DESKew <Deskew>
Sets a delay for the selected channel.
Deskew compensates delay differences between channels caused by the different
length of cables, probes, and other sources.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Deskew> Deskew value
Range: -100E-9 to 100E-9
Increment: 800E-12
*RST: 0
Default unit: s
Waveform Setup