Remote Control Commands
R&S
®
Scope Rider RTH
304User Manual 1326.1578.02 ─ 15
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<State> ON | OFF
*RST: OFF
CHANnel<m>:SCALe <Scale>
Sets the vertical scale (vertical sensitivity) of the indicated waveform.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Scale> Scale value, given in Volts per division
Range: 2E-3 to 100
Increment: 1E-3
*RST: 0.05
Default unit: V/div
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 8 vertical divisions of the diagram. Use the com-
mand alternatively to CHANnel<m>:SCALe.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Range> Voltage range value
Range: 200E-9 to 8E+6
Increment: 10E-6
*RST: 0.4
Default unit: V
CHANnel<m>:PROBe <ProbeSettings>
Sets the attenuation factor of the connected probe. The vertical scaling is adjusted
accordingly, and measured values are multiplied by this factor so that the displayed
values are equal to the actual signal values.
Make sure to set the attenuation factor on the instrument according to the probe being
used. Otherwise, the measurement results do not reflect the actual voltage level, and
you might misjudge the actual risk.
If AC coupling is set, the attenuation of passive probes has no effect, and voltage is
applied to the instrument with factor 1:1. Observe the voltage limits, otherwise you can
damage the instrument.
Waveform Setup