Instructions

Table Of Contents
Network Connections
R&S
®
Scope Rider RTH
299User Manual 1326.1578.02 ─ 15
To use the instrument as access point
1.
Press the key.
2. Tap "Wireless LAN".
3. Only for option R&S RTH-K200: Select the "Country" where you use the instru-
ment.
The data sheet contains list of countries where the option R&S RTH-K200 is certi-
fied or permitted.
4. Select "Wireless Mode" = "Access Point" in the "Wireless LAN" menu.
5. If you connect for the first time, change the default "Passphrase". You can also
change the identifier of the instrument, the "SSID".
6. Enable "Wireless State".
7. On your remote device, set up the connection to the instrument. Select the SSID,
and enter the passphrase.
The detailed procedure is described in the documentation of your device.
To use the instrument as client
1.
Press the key.
2. Tap "Wireless LAN".
3. Only for option R&S RTH-K200: Select the "Country" where you use the instru-
ment.
The data sheet contains list of countries where the option R&S RTH-K200 is certi-
fied or permitted.
4. Select "Wireless Mode" = "Client" in the "Wireless LAN" menu.
5. Enable "Wireless State".
The "Choose Access Point" window opens.
6. Tap the "SSID" of the required router.
7. Enter the password of the required router in "Passphrase".
The connection status is shown in "Status"
Wireless LAN Connection (Option R&S RTH-K200/200US)