Frequency Counter (R&S RTH-K33)
R&S
®
Scope Rider RTH
266User Manual 1326.1578.02 ─ 15
Reference
Selects the reference to be used for the counter.
"Counter 1 / 2"
The measured value of the other counter is used as a reference.
Specify the nominal "Reference Frequency" for counter 2.
The deviation of the nominal frequency to the measured reference
frequency is internally removed from the measured frequency of
counter 1.
"Internal"
An internal reference is used to determine the counter frequency.
Remote command:
COUNter<m>:SENSe:REFerence:STATe on page 415
COUNter<m>:SENSe:REFerence:VALue on page 416
Restart
Resets all statistical values and starts a new counter measurement.
Remote command:
COUNter<m>:CALCulate:AVERage:CLEar on page 411
COUNter<m>:INITiate on page 414
Counter Settings