Instructions
Table Of Contents
- Cover
- Contents
- 1 For Your Safety
- 2 Getting Started
- 3 Waveform Setup
- 3.1 Connecting Probes
- 3.2 Vertical Setup
- 3.3 Horizontal Setup
- 3.4 Acquisition Control
- 3.5 Roll Mode
- 3.6 Trigger
- 3.6.1 General Trigger Settings
- 3.6.2 Edge Trigger
- 3.6.3 Glitch Trigger
- 3.6.4 Width Trigger
- 3.6.5 Video/TV Trigger
- 3.6.6 External Trigger (R&S RTH1002)
- 3.6.7 Pattern Trigger (R&S RTH-K19)
- 3.6.8 State Trigger (R&S RTH-K19)
- 3.6.9 Runt Trigger (R&S RTH-K19)
- 3.6.10 Slew Rate Trigger (R&S RTH-K19)
- 3.6.11 Data2Clock Trigger (R&S RTH-K19)
- 3.6.12 Serial Pattern Trigger (R&S RTH-K19)
- 3.6.13 Timeout Trigger (R&S RTH-K19)
- 3.6.14 Interval Trigger (R&S RTH-K19)
- 3.6.15 Window Trigger (R&S RTH-K19)
- 3.6.16 Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)
- 4 Waveform Analysis
- 5 Mask Testing
- 6 Spectrum Analysis
- 6.1 FFT Mode
- 6.2 Spectrum Mode (Option R&S RTH-K18)
- 6.3 Harmonics Measurement (Option R&S RTH-K34)
- 7 Multimeter Measurements
- 8 Data Logging
- 9 Protocol Analysis
- 10 Logic Analyzer (R&S RTH-B1 MSO)
- 11 Frequency Counter (R&S RTH-K33)
- 12 Documenting Results
- 13 General Instrument Setup
- 14 Network Connections
- 15 Remote Control Commands
- 15.1 Conventions used in Command Description
- 15.2 Mode
- 15.3 Waveform Setup
- 15.3.1 Automatic Setup
- 15.3.2 Vertical Setup
- 15.3.3 Horizontal Setup
- 15.3.4 Acquisition Control
- 15.3.5 Trigger
- 15.3.5.1 General Trigger Settings
- 15.3.5.2 Edge Trigger
- 15.3.5.3 Glitch Trigger
- 15.3.5.4 Width Trigger
- 15.3.5.5 Video/TV Trigger
- 15.3.5.6 External Trigger (R&S RTH1002)
- 15.3.5.7 Pattern Trigger (R&S RTH-K19)
- 15.3.5.8 State Trigger (R&S RTH-K19)
- 15.3.5.9 Runt Trigger (R&S RTH-K19)
- 15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
- 15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
- 15.3.5.12 Serial Pattern Trigger (R&S RTH-K19)
- 15.3.5.13 Timeout Trigger (R&S RTH-K19)
- 15.3.5.14 Interval Trigger (R&S RTH-K19)
- 15.3.5.15 Window Trigger (R&S RTH-K19)
- 15.4 Waveform Analysis
- 15.5 Mask Testing
- 15.6 Spectrum Analysis
- 15.7 Digital Multimeter (R&S RTH1002)
- 15.8 Voltmeter (R&S RTH1004)
- 15.9 Counter Mode (R&S RTH-K33)
- 15.10 Data Logging
- 15.11 Protocol Analysis
- 15.12 Logic Analyzer (R&S RTH-B1 MSO)
- 15.13 Documenting Results
- 15.14 General Instrument Setup
- 15.15 WLAN Connection (Option R&S RTH-K200/200US)
- 15.16 User Scripting (R&S RTH-K38)
- Annex
- List of Commands
Protocol Analysis
R&S
®
Scope Rider RTH
238User Manual 1326.1578.02 ─ 15
Identifer Relation
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Remote command:
TRIGger:SENT:ICONdition on page 483
Sync Pulse Error
Detects a synchronization/calibration pulse error in transmission sequences of the fast
channel. An error occurs when:
●
The duration of the synchronization/calibration pulse (in ticks) is less than 56*(1-
clock tolerance) or more than 56*(1+clock tolerance).
●
The synchronization/calibration pulse duration of frame (n–1) varies by more than
1.5625% from the calibration/sync pulse duration of frame (n).
Remote command:
TRIGger:SENT:PULSeerror on page 484
Pulse Period Error
Detects an error in the calibration/sync pulse in transmission sequences of the fast
channel. An error occurs when a nibble has any of the following:
●
Number of ticks at low is fewer than 4 ticks.
●
Nibble value< 0 (fewer than 12 ticks) or > 15 (more than 27 ticks).
Through the pulse period error, you can also identify a nibble length error of the previ-
ous transmission sequence.
Remote command:
TRIGger:SENT:PPERioderror on page 484
Fast CRC Error
Detects a checksum error in the transmission sequences of the fast channel. The CRC
length is 4 bits.
Remote command:
TRIGger:SENT:FCRCerror on page 483
Slow CRC Error
Detects a checksum error in serial messages of the slow channel. The CRC length is 4
bits for short serial messages, and 6 bit of enhanced serial messages.
Remote command:
TRIGger:SENT:SCRCerror on page 484
Frame Length Error
Detects frame length errors in transmission sequences when Pause Pulse is set to
"Constant frame length".
A frame length error occurs, when the total length of the transmission sequence
(including pause pulse) does not match the Frame Length.
Remote command:
TRIGger:SENT:IRFLength on page 484