Scope Rider RTH
223User Manual 1326.1578.02 ─ 15
To set a user-defined value, double-tap the "Bit Rate" field, then enter the value and
unit using the displayed keypad. The "Predefined Bit Rates" setting is automatically set
to "User Bit Rate".
Remote command:
BUS:LIN:BITRate on page 470
Threshold, Technology, Find Level
Sets the threshold value for digitization of signals. If the signal value is higher than the
threshold, the signal state is high. Otherwise, the signal state is considered low.
You can select the threshold voltages for various technologies in the "Technology" list,
or set a user-defined value in "Threshold". You can also let the instrument set the
appropriate threshold using "Find Level".
Changing the thresholds in the bus configuration changes also the thresholds of ana-
log channels in the "Vertical" menu, or the thresholds of logic channels in the "Logic"
menu.
Remote command:
BUS:LIN:TECHnology on page 471
BUS:LIN:DATA:THReshold on page 470
BUS:SETReflevels on page 431
CHANnel<m>:THReshold:FINDlevel on page 309
9.6.3 LIN Trigger Settings
Access: [Setup] ([Trigger]) > "Trigger type" = "Bus"
LIN Trigger.................................................................................................................. 224
Checksum Error.......................................................................................................... 224
Parity Error.................................................................................................................. 224
Sync Error................................................................................................................... 224
Identifier Pattern..........................................................................................................224
Identifier Relation........................................................................................................ 225
LIN (Option R&S
RTH-K3)