Instructions

Table Of Contents
Protocol Analysis
R&S
®
Scope Rider RTH
213User Manual 1326.1578.02 ─ 15
See also Chapter 3.6.12.1, "Pattern Definition", on page 71.
Remote command:
TRIGger:CAN:IDENtifier on page 460
Identifier Relation
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Remote command:
TRIGger:CAN:ICONdition on page 460
FDF Bit
The bit determines whether a frame is CAN or CAN-FD. It corresponds to the EDL bit
(extended data length), which only exists in CAN FD format. If you do not know if the
signal is CAN or CAN FD, you can use this bit to identify the format.
Remote command:
TRIGger:CAN:FDATa:FDF on page 462
BRS Bit
The setting is available in CAN FD option R&S RTH-K9.
Sets the the bit rate switch bit.
Value 1 means that the bit rate switches from the "Arbitration rate" to the faster "Data
rate".
Remote command:
TRIGger:CAN:FDATa:BRS on page 461
ESI Bit
The setting is available in CAN FD option R&S RTH-K9.
Sets the error state indicator bit. If set to "Dominant", the bit indicates an error active
state.
Remote command:
TRIGger:CAN:FDATa:ESI on page 462
Data Pattern
Specifies the data pattern to be found, in binary or hex format. Enter the pattern in
MSB first bit order.
See also Chapter 3.6.12.1, "Pattern Definition", on page 71.
Remote command:
TRIGger:CAN:DATA on page 459
Data Relation
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Remote command:
TRIGger:CAN:DCONdition on page 459
CAN and CAN FD (Options R&S
RTH-K3, R&S RTH-K9)