Source: CS, Clock, MOSI, MISO
Set the input channels of the SPI lines.
If option R&S RTH-B1 is installed, digital channels can also be used as source.
Remote command:
BUS:SPI:SSEL:SOURce on page 441
BUS:SPI:SCLK:SOURce on page 441
BUS:SPI:MISO:SOURce on page 441
BUS:SPI:MOSI:SOURce on page 441
Polarity
Selects whether the chip select signal is high active (high = 1) or low active (low = 1).
Remote command:
BUS:SPI:SSEL:POLarity on page 441
Slope
Selects if data are sampled on the rising or falling slope of the clock. The clock slope
marks the begin of a new bit.
Remote command:
BUS:SPI:SCLK:SLOPe on page 441
Thresholds, Technology, Find Level
Sets the threshold value for digitization of signals. If the signal value is higher than the
threshold, the signal state is high. Otherwise, the signal state is considered low.
You can select the threshold voltage on all lines for various types of integrated circuits
in the "Technology" list, or set a user-defined value for each line in "Thresholds". You
can also let the instrument set the appropriate thresholds using "Find Level".
Changing the thresholds in the bus configuration changes also the thresholds of ana-
log channels in the "Vertical" menu, or the thresholds of logic channels in the "Logic"
menu.
SPI (Option R&S