2.3.2 Top View
The R&S RTH1002 has two BNC input connectors CH1 and CH2, and two 4 mm
banana plug inputs for multimeter measurements. The channel inputs have double
channel-to-channel isolation that allows for independent floating measurements at
each input. The DMM input is fully isolated from scope inputs, interfaces, and ground.
Figure 2-4: Top view of R&S
RTH1002
The R&S RTH1004 has four BNC input connectors CH1, CH2, CH3, CH4. The channel
inputs have double channel-to-channel isolation that allows for independent floating
measurements at each input.
Figure 2-5: Top view of R&S
RTH1004
Shock hazard caused by high voltages
To avoid electrical shock and personal injury, and to prevent damage to the instrument
or any other products connected to it, observe the following:
Do not apply input voltages above the rating of the instrument and the accessories.
Use only probes, test leads, and adapters that comply with the measurement cate-
gory (CAT) of your measurement task.
Test leads and measurement accessories used for multimeter measurements on a
live mains circuit must be rated for CAT III or CAT IV according to IEC 61010-031.
The voltage of the measured circuit must not exceed the rated voltage value.
Maximum input voltage:
At BNC inputs: CAT IV 300 V
With probe R&S RT-ZI10 or R&S RT-ZI11: CAT IV 600 V, CAT III 1000 V
Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings: V RMS (50 to 60 Hz) for AC sine wave and V DC for DC applications.