You can also use the label list to trigger on an identifier or address. Instead of entering
the value, you select the name, which is defined in the label list.
Available file formats are PTT, CSV, and DBC (CAN only).
Label lists are protocol-specific. Their contents are described in the corresponding pro-
tocol chapters:
Chapter 9.2.4, "I²C Label List", on page 187
Chapter 9.5.4, "CAN Label List", on page 215
Chapter 9.6.4, "LIN Label List", on page 225
Chapter 9.7.4, "SENT Label List", on page 239
9.1.3.1 Using Label Lists
To load a label list and display the labels:
1. Save the label list file on a USB flash drive, or on the microSD card.
2. Press and hold the [BUS] key until the menu opens.
3. Configure the protocol.
4. In the "Bus" menu, tap "Load Label List".
5. Navigate to the label list file, select it, and tap "Select".
6. To read the label list, tap "Show Label List".
7. To use the labels in the display of the decoded data, tap "Display Labels".
To trigger on an identifier or address using the label:
Prerequisites: The bus is configured, decoding is enabled, and a decoded signal is visi-
ble.
1. Open the "Trigger" menu.
2. Set the following trigger settings:
"Trigger Mode" = "Normal"
"Trigger Type" = "Bus"
"<Protocol> Trigger" = "Identifier", or "Identifier + Data", or "Address", or
"Address and Data".
3. Tap "Identifier from Label" or "Address from Label".
4. Select the label. Tap "Back".
If the selected identifier or address is available in the acquired data, its value is
shown in the menu, and the instrument triggers on it.