Instructions

Table Of Contents
Data Logging
R&S
®
Scope Rider RTH
161User Manual 1326.1578.02 ─ 15
Figure 8-2: One-log-view with highligted scaling mode
To change the vertical scale and position manually, press the vertical [RANGE] and
[POS] keys.
In both log views (one-log and all-log), the keys affect only the selected channel.
To return a single logger channel to autoscale:
a) Switch to the one-log-view of the channel.
b) Press [AUTOSET].
To return a all logger channels to autoscale:
a) Switch to the all-log-view.
b) Press [AUTOSET].
Note that [AUTOSET] sets also the horizontal scaling to "auto".
Remote commands:
LOGGer:AUToset on page 420
LOGGer:MEASurement<m>:VERTical:AUTO on page 420
LOGGer:MEASurement<m>:VERTical:UPPer on page 420
LOGGer:MEASurement<m>:VERTical:LOWer on page 420
LOGGer:MEASurement<m>:VERTical:DEViation on page 421
LOGGer:MEASurement<m>:VERTical:MEAN on page 421
Using the Logger