Multimeter Measurements
R&S
®
Scope Rider RTH
156User Manual 1326.1578.02 ─ 15
Remote command:
Chapter 15.8.3, "Relative Measurements", on page 403
Reference/Reference Manual
"Reference" activates or deactivates the manual reference value.
If active, enter the reference value in "Reference Manual".
Restart Test
Restarts the measurement and also restarts statistical calculation.
[RANGE] / [POS]
In meter mode, the vertical [RANGE] and [POS] keys adjust the measurement range of
the selected channel.
Remote command:
Chapter 15.8.2, "Measurement Configuration", on page 399
7.2.5 Vertical Settings
In "Meter" mode, only a few vertical channel settings are relevant. You can also config-
ure the vertical settings for the input signal in "Scope" mode. When you select the
"Meter" mode, the configuration is maintained.
Channel Index
Selects the channel to be configured. All settings in the vertical menu belong to the
selected channel.
You can also press and hold the channel key to open the corresponding vertical menu.
Probe Setting
Selects the attenuation factor of the connected probe.
The menu lists the common attenuation factors. If the required factor is not in the list,
select "User" and set the "Probe Factor" on page 157.
Make sure to set the attenuation factor on the instrument according to the probe being
used. Otherwise, the measurement results do not reflect the actual voltage level, and
you might misjudge the actual risk.
Voltmeter (R&S
RTH1004)