Instructions

Table Of Contents
Spectrum Analysis
R&S
®
Scope Rider RTH
144User Manual 1326.1578.02 ─ 15
<Order of harmonic>, <limit value in % of fundamental>
Comments introduced by "#" are allowed.
If a limit is not provided for a harmonic, the harmonic is displayed and its results
are included in the export, but not in the limit check.
The fundamental is always displayed and need not be included in the list.
Example:
RTH,Version,1.0
Harmonics,Limit[%]
2,10.5
4,0.004
20,3.7
10
13,0.01
3,0.01
6.3.6.2 Export File Format for Harmonic Results
The harmonic results export file format is similar to a waveform export file; it is also
stored in CSV format. You can convert the comma-separated text to columns (see
Chapter 12.4.3.3, "Converting CSV to Excel Files", on page 278).
The harmonics results file contains some general header data and then a set of result
data for each harmonic. A maximum of 64 results can be provided, but the actual num-
ber of harmonics may be restricted (see "Limits, Loaded File" on page 141). Each set
of data contains the results for a specific harmonic in each input channel.
Table 6-7: Header data (converted to table)
Harmonic Results
[Description - not included in file]
Model RTH1004
SerialNumber xxxxx
Firmware Version '1.50'
ID CH1 CH2 CH3 CH4
Results for each channel
NbOfResults [n]
Number of results included in averaging and statis-
tics operations
RMS Unit
Unit used for RMS results
RMS
Current RMS of the signal power (
RMS Max
Maximum RMS of the signal power
RMS Min
Minimum RMS of the signal power
THDf/ THDr [%]
Current THD in relation to the RMS amplitude of the
fundamental frequency (THDf) or the total power of
the signal (THD
R
)
Harmonics Measurement (Option R&S RTH-K34)