Instructions
Table Of Contents
- Cover
- Contents
- 1 For Your Safety
- 2 Getting Started
- 3 Waveform Setup
- 3.1 Connecting Probes
- 3.2 Vertical Setup
- 3.3 Horizontal Setup
- 3.4 Acquisition Control
- 3.5 Roll Mode
- 3.6 Trigger
- 3.6.1 General Trigger Settings
- 3.6.2 Edge Trigger
- 3.6.3 Glitch Trigger
- 3.6.4 Width Trigger
- 3.6.5 Video/TV Trigger
- 3.6.6 External Trigger (R&S RTH1002)
- 3.6.7 Pattern Trigger (R&S RTH-K19)
- 3.6.8 State Trigger (R&S RTH-K19)
- 3.6.9 Runt Trigger (R&S RTH-K19)
- 3.6.10 Slew Rate Trigger (R&S RTH-K19)
- 3.6.11 Data2Clock Trigger (R&S RTH-K19)
- 3.6.12 Serial Pattern Trigger (R&S RTH-K19)
- 3.6.13 Timeout Trigger (R&S RTH-K19)
- 3.6.14 Interval Trigger (R&S RTH-K19)
- 3.6.15 Window Trigger (R&S RTH-K19)
- 3.6.16 Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)
- 4 Waveform Analysis
- 5 Mask Testing
- 6 Spectrum Analysis
- 6.1 FFT Mode
- 6.2 Spectrum Mode (Option R&S RTH-K18)
- 6.3 Harmonics Measurement (Option R&S RTH-K34)
- 7 Multimeter Measurements
- 8 Data Logging
- 9 Protocol Analysis
- 10 Logic Analyzer (R&S RTH-B1 MSO)
- 11 Frequency Counter (R&S RTH-K33)
- 12 Documenting Results
- 13 General Instrument Setup
- 14 Network Connections
- 15 Remote Control Commands
- 15.1 Conventions used in Command Description
- 15.2 Mode
- 15.3 Waveform Setup
- 15.3.1 Automatic Setup
- 15.3.2 Vertical Setup
- 15.3.3 Horizontal Setup
- 15.3.4 Acquisition Control
- 15.3.5 Trigger
- 15.3.5.1 General Trigger Settings
- 15.3.5.2 Edge Trigger
- 15.3.5.3 Glitch Trigger
- 15.3.5.4 Width Trigger
- 15.3.5.5 Video/TV Trigger
- 15.3.5.6 External Trigger (R&S RTH1002)
- 15.3.5.7 Pattern Trigger (R&S RTH-K19)
- 15.3.5.8 State Trigger (R&S RTH-K19)
- 15.3.5.9 Runt Trigger (R&S RTH-K19)
- 15.3.5.10 Slew Rate Trigger (R&S RTH-K19)
- 15.3.5.11 Data2Clock Trigger (R&S RTH-K19)
- 15.3.5.12 Serial Pattern Trigger (R&S RTH-K19)
- 15.3.5.13 Timeout Trigger (R&S RTH-K19)
- 15.3.5.14 Interval Trigger (R&S RTH-K19)
- 15.3.5.15 Window Trigger (R&S RTH-K19)
- 15.4 Waveform Analysis
- 15.5 Mask Testing
- 15.6 Spectrum Analysis
- 15.7 Digital Multimeter (R&S RTH1002)
- 15.8 Voltmeter (R&S RTH1004)
- 15.9 Counter Mode (R&S RTH-K33)
- 15.10 Data Logging
- 15.11 Protocol Analysis
- 15.12 Logic Analyzer (R&S RTH-B1 MSO)
- 15.13 Documenting Results
- 15.14 General Instrument Setup
- 15.15 WLAN Connection (Option R&S RTH-K200/200US)
- 15.16 User Scripting (R&S RTH-K38)
- Annex
- List of Commands
Spectrum Analysis
R&S
®
Scope Rider RTH
140User Manual 1326.1578.02 ─ 15
Limits, Loaded File...................................................................................................... 141
Open........................................................................................................................... 141
Display Harmonics, Max No Harmonics......................................................................142
Average.......................................................................................................................142
[CH <n>]
Selects the channel for which the captured data is analyzed in "Harmonic" mode. Har-
monic measurements can be performed on all channels simultaneously. The results
are indicated by different colored bars and numbers, corresponding to the channel but-
ton color.
Remote command:
CHANnel<m>:STATe on page 303
Scale
Switches the scale for the measured harmonic levels between logarithmic (dB) and lin-
ear (percent). The values are relative to the level measured for the fundamental fre-
quency.
Remote command:
HARMonic:SCALe on page 374
THD Type
Switches between different calculation methods for the total harmonic distortion.
"THD
F
"
The RMS amplitude (voltage or current) of the harmonics relative to
the RMS amplitude of the fundamental component
"THD
R
"
The RMS amplitude of the harmonics relative to the RMS amplitude
of the input signal
Remote command:
HARMonic:THDType on page 375
Statistic
By default, the numeric results indicate the currently measured values. Optionally, you
can switch to the minimum or maximum values. Which value is calculated is indicated
in the result display (see "(9) Result type for numeric results" on page 136).
Note: The bargraph always displays both the current and the maximum values.
Remote command:
HARMonic:STATistic on page 375
Clear Results
Clears the results for previous acquisitions used in statistical evaluation (see "Statistic"
on page 140 and "Average" on page 142).
Remote command:
HARMonic:CLEar on page 371
Fundamental Frequency, Value
Defines the basis of the harmonics measurement. Harmonics are determined as multi-
ples of this frequency.
For a user-defined frequency, select "User" and enter the "Value".
Harmonics Measurement (Option R&S
RTH-K34)