Instructions

Table Of Contents
Mask Testing
R&S
®
Scope Rider RTH
106User Manual 1326.1578.02 ─ 15
Note: The acquisition mode "Envelope" is not available while running mask tests.
2. If you want to test a mathematical (math) waveform, generate a math waveform,
see Chapter 4.4, "Mathematics", on page 91.
3. Select the "Mask" menu.
4. Select the mask channel that is assigned to the tested waveform.
5. To activate the mask, tap "State".
Note: If the selected mask was used before, the stored mask is displayed.
6. To set the mask width, enter "Width X" and "Width Y".
7. Select the "Action on Violation".
This action is applied to all masks.
8. To create the mask, tap "Create Mask".
9. To start and stop mask testing, press the [RUN STOP] key.
Analyzing the mask test
To view the details of the waveform at a mask test violation, you can use the zoom
function, see Chapter 4.1, "Zoom", on page 78. Zooming is only possible while the
mask test is stopped.
5.4 Mask Settings
Access: "Mask" menu
Mask Settings