Manual

Table Of Contents
Trusted
TM
Valve Manager T8031
Issue 2 Oct 05 PD-T8031 39
4.9.3. Profile Editor
The Profile Editor dialog box allows you to configure information that the Valve Manager will use
when initiating a test and applying additional pass/fail criteria. You must select the intendedGoal
State for the valve and supply aMaximum Test Time. TheGoal State field corresponds to the
_EST variable in the application and is used to indicate the desired valve position when a test is
i
nitiated. TheMaximum Test Time field corresponds to the_ETM variable which specifies the
maximum duration of a valve test. See the T8449 Product Description for a complete description of
how these variables are used.
TheThreshold Settings section of the dialog allows you to specify additional pass/fail criteria to be
applied to the state transition times following a test. The possible valve positions (as specified in the
State Order dialog) are displayed to left, along with some very basic statistics: Minimum, Maximum,
Average, and Count. Theses statistics are calculated from all previously recorded transition time
information. A value of 0 (zero) for either theMin orMax threshold value will cause the Valve
manger to ignore that threshold setting. Usually, the threshold settings are set to 0 (zero) initially.
After one or more tests have been completed, the threshold settings are then updated.
Click the “Finish” button to save the profile once the information is complete.
Figure 38 Profile Editor