User Manual

14
Connections
Connect the output circuit as shown on page one with the
multimeter in place of the receiving device.
Connect the 931U-C9A2C-OP up to a PC running the setup/
Calibration software (as if you were going to change the setup).
Input Calibration points
Range Values Notes
Volts/Thermocouple 10 mV, 20 mV,
50 mV, 80 mV,
200 mV, 1 V
If the internal switches are set to 10 V, you must
supply a signal ten times the displayed value, e.g. 10
V for the 1V point.
mA 20 mA
RTD/Resistance
100 Ω, 200 Ω,
500 Ω, 800 Ω,
2 kΩ, 10 kΩ
Use four wire mode
Individual calibration As selected Allows you to recalibrate any of the points above
Fine adjust Current input Fine tune the adjustment for the range you have
selected.
All All ranges Complete calibration of all points. Internal switches
must be in the 1V position.