User guide

Installation
8
3. 3 Installation Instructions
Fig. 3: Typical applications of the Bulletin 836E to measure pressure in gases, steam and liquids.
Bulletin 836E pressure switch
Shut-off assembly
U-pipe
For typical applications of the Bulletin 836E, see Figure 3 above:
The image on the left illustrates how the 836E should be applied for pressure measurement in
gases. The switch should be mounted with a shut-off assembly above the sampling nozzle so
that any condensate can drain off into the process.
The center image shows the correct installation for pressure measurement in steam. Note how
the Bulletin 836E is mounted with a U-pipe below the sampling nozzle. Fill the U-pipe with fill
fluid before commissioning.
For pressure measurement in liquids, see the drawing on the right of Figure 3 above. The
Bulletin 836E should be mounted below or at the same level as the sampling nozzle.
Additional considerations:
Do not mount the device in the product flow or at a point where it could be affected by
pressure pulses.
Calibration and testing are easier if the device is mounted downstream of a shut-off assembly.
The orientation of the Bulletin 836E can result in zero point shift, i.e. the measured value does
not display zero in a non-pressurized state. This zero point shift can be corrected (see Section4
– Operation).
The digital display can be electronically rotated 180° (see Section 4 – Operation).
The housing can be rotated up to 310° for optimal readability and ease of wiring.
➂➂
Gas
Steam
Liquid