User guide
Operation
17
4 - 20 Analog output
Value for 4 mA
(LRV)
Enter lower range
value in increments
of 0.1%
Value for 20 mA
(URV)
Enter upper range
value in increments
of 0.1%
Pressure applied for
4 mA (LRV)
Use measured
pressure as lower
range value
Pressure applied for
20 mA (URV)
Use measured
pressure as upper
range value
Error current Current value in
event of error:
MIN = ≤ 3.6 mA
MAX = ≥ 21.0 mA
HOLD = last value
Range of adjustment: LRL = Lower Range Limit; URL = Upper Range Limit;
LRV = Lower Range Value; URV = Upper Range Value
Turndown up to 4:1, LRV must be lower than URV