Manual

Publication 1762-UM002A-EN-P - July 2002
Specifications A-19
Figure A.15 Module Accuracy at 25°C (77°F) Ambient for Type S Thermocouple
Using 10, 50, and 60 Hz Filter
0. 0
0. 5
1. 0
1. 5
2. 0
2. 5
0 200 400 600 800 1000 1200 1400 1600 1800
10 Hz
50 Hz
60 Hz
0. 0
0. 5
1. 0
1. 5
2. 0
2. 5
3. 0
3. 5
4. 0
4. 5
0 500 1000 1500 2000 2500 3000 3500
10 Hz
50 Hz
60 Hz
Thermocouple Temperature °C
Thermocouple Temperature °F
Accuracy °C Accuracy °F