Setting Up Measurements Chapter 2
Typical Magnitude and Phase Measurement Definitions
this is what the Measurement Definition pane for one through four orders might
look like. All measurement definitions have the same units and collection
specification.
Figure 15 - Magnitude and Phase Measurement Definitions
The following table shows the collection specification for Mag & Phase.
The transducer specification in the Mag & Phase collection specification is
shown in the table below.
The filter specifications are shown in the table below.
Table 20 - Collection Specification for Mag & Phase Measurement Definitions
Transducer Window Signal Detection Fmax Lines Phase Order Norm? Averages
Accelerometer Hanning Peak 500 Hz 400 Yes Yes 4 linear
Table 21 - Transducer Specification in Mag & Phase Collection Specification
Name Base Unit Input Type Units Calibration Value DC Offset
Accelerometer Acceleration ICP Accel g 100 0
Table 22 - Filter Specifications in Mag & Phase Measurement Definitions
Name Low Cutoff High Cutoff Data Collector Filter Type
1st Order 0.8 Orders 1.2 Orders None
2nd Order 1.8 Orders 2.2 Orders None
3rd Order 2.8 Orders 3.2 Orders None
4th Order 3.8 Orders 4.2 Orders None