144 Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012
Chapter 4 Collecting and Reviewing Data
Create a User-defined Point
Follow these steps to create a user-defined point.
1. Select Data Collection from the Main Menu and press the Enter key.
The Data Collection screen appears.
2. Select the route and press Enter.
The Enpac 2500 data collector always associates Offroute measurements
with a list.
3. Press F2 (Offroute) to collect an Offroute measurement.
gSE Performs a standard vibration Acceleration measurement and collects waveform data.
These are the settings for the measurement parameters.
Table 36 - Types of Offroute Measurements
Type Description
Parameter Value
Units Enveloped Acceleration
Display Format Spectrum
Band High-pass frequency gSE BPF 5 kHz
Fmax 10kHz
Lines/Samples 400
Averages 4
Input Channel X
Overlap 50%
Ext. Trigger Slope Positive
Trigger Level 2 V
Bin Zeroing On
Typ e Spec tra l