The data collector takes vibration readings, calculates new trim weight(s)
and displays the values.
If the rotor is still out of balance, repeat adding the trim weight and
taking residual measurements until the rotor is properly balanced.
If several trim correction runs have been performed (resulting in several
trim weights), values may be combined into one permanent trim
weight.
If the balance never improves, check to see if you entered the correct
trial weight size and location, or check other factors such as
misalignment or a faulty bearing.
However, if the vibration has been reduced to an acceptable level,
balancing is complete.
9. Save your balancing data, if desired, by pressing F3 (Memory).
See Save a Balance Measurement on page 61