User guide

10 Rockwell Automation Publication GMSI10-IN026C-EN-P - February 2014
XM-160, XM-161, and XM-162 Direct Vibration Modules
Emissions
CISPR 11 (IEC 61000-6-4)
Class A
ESD immunity
IEC 61000-4-2
8 kV air discharges
Radiated RF immunity
IEC 61000-4-3
10V/m with 1 kHz sine-wave 80% AM from 80…2000 MHz
10V/m with 200 Hz 50% Pulse 100% AM at 900 MHz
10V/m with 200 Hz 50% Pulse 100% AM at 1890 MHz
3V/m with 1 kHz sine-wave 80% AM from 2000…2700 MHz
EFT/B immunity
IEC 61000-4-4
±2 kV at 5 kHz on shielded power ports
±2 kV at 5 kHz on shielded signal ports
±2 kV at 5 kHz on XMbus port
Surge transient immunity
IEC 61000-4-5
±2 kV line-earth(CM) on shielded power ports
±2kV line-earth(CM) on shielded signal ports
±2kV line-earth(CM) on XMbus port
Conducted RF immunity
IEC 61000-4-6
10V rms with 1 kHz sine-wave 80% AM from 150 kHz…80 MHz
Enclosure type rating None (open-style)
Voltagesranges XM-160/XM-162 Supply:
24V DC, 0.19 A
XM-161 Supply:
24V DC, 0.31 A
Power dissipation XM160/XM-162: 4.6 W @ 24V DC (4.3 W @ 18V DC, 4.9 W @ 32V DC), Class 2/SELV
XM-161: 7.4 W @ 24V DC (7 W @ 18V DC, 8 W @ 32V DC), Class 2/SELV
External over-current protection 440-5AFUSEKIT for SELV/PELV power source
Isolation voltage Not rated
Wiring category
(1)
2 - on shielded power and shielded signal ports
3 - on Serial ports
2 - on XMbus ports
Wire type Power and signal connections: shielded
North American temp code T4
IEC temp code T4
(1) Use this Conductor Category information for planning conductor routing. Refer to Industrial Automation Wiring and Grounding Guidelines, publication 1770-4.1.
Attribute XM-16X