490 List of flags
Roche Diagnostics
D
igitalLightCycler
®
System · Software version 1.0 · User Assistance · Publication version 1.1
Flag code Severity Description Actions
IP003
Informational No nanowell columns detected within the
expected overlap area during stitching
Restart the analyzer. If the flag occurs
repeatedly, contact your Roche Service
representative.
u Restarting the analyzer (451)
IP004
Critical Data integrity check failed
Restart the analyzer. If the flag occurs
repeatedly, contact your Roche Service
representative.
u Restarting the analyzer (451)
PE001
Critical Partitioning engine: Lane partitioning
failed
Restart the partitioning engine. If the flag
occurs repeatedly, contact your Roche Service
representative.
u Restarting the analyzer (451)
PE002
Informational Partitioning engine: Lane partitioning
data missing
Restart the partitioning engine. If the flag
occurs repeatedly, contact your Roche Service
representative.
u Restarting the analyzer (451)
PE003
Critical Partitioning engine: Empty lane
RA000
Critical Unhandled exception in results
interpretation.
Restart the analyzer. If the flag occurs
repeatedly, contact your Roche Service
representative.
u Restarting the analyzer (451)
RA001
Informational Upper limit of quantification not met.
RA002
Informational Lower limit of quantification not met.
RA003
Critical Positive control concentration is too high.
RA004
Critical Positive control concentration is too low.
RA005
Critical Negative control concentration is too
high.
RA006
Critical Internal control concentration is too high.
RA007
Critical Internal control concentration is too low.
RA008
Critical Too many invalid nanowells
If the flag occurs repeatedly, contact your
Roche Service representative.
RA009
Critical Integrity check failed.
If the flag occurs repeatedly, contact your
Roche Service representative.
RA011
Critical Too many artifact nanowells
Ensure there are no scratches, finger prints,
dust, or other artifacts on the nanowell plate
lanes that compromise the image integrity.
If the flag occurs repeatedly, contact your
Roche Service representative.
RA012
Critical Merging verification failed.
If the flag occurs repeatedly, contact your
Roche Service representative.
RA013
Critical Copy number control value is too high.
RA014
Critical Copy number control value is too low.
RA015
Critical TOI concentration is above upper limit of
validity.
RA016
Critical TOI concentration is below upper limit of
validity.
y Table of flags
18 Flags