User`s manual
The meter is in the “memory”
function, allowing you to
recall 7, 14, and 30-day
averages.
This blood glucose result was
marked with a result marker
flag (asterisk) (see page 16).
Do not use individual values stored in
memory to make therapy adjustments.
Temperature is above or below
the operating range of test
strips (14°to 40°C).
The result you have obtained
may not be accurate. Move to
an area between 14° and
40°C, wait 5 minutes, and repeat the test.Do not
artificially heat or cool the meter.
Memory Displays
The meter is in the “memory”
function, allowing you to recall
results from memory
(see page 20).
If you are reviewing memory
and a result appears with a
temperature symbol as
shown, this result may
not be accurate because
the test was performed outside
the operating range of the test
strips (14° to 40°C ).