Storage and Handling
Keep your meter free of dust. Protect it from extreme
temperature and humidity, and handle it with care.If
meter is dropped, perform a quality control check with
the glucose control solutions before doing a blood
glucose test.
The travel case is designed to let you store a variety of
supplies you may need and helps to protect your meter.
Caring For Your Meter
Display Check
This verifies that all parts of
the display are working. Make
sure the meter is off. Press and
hold down the ON/OFF button.
The display should look exactly
like the picture at right. If it
does not, call our Diabetes Care Info-Line.
Cleaning
Clean the outside of the meter,as needed, using a soft
cloth that has been slightly dampened with one of the
following:
70% alcohol
•a solution of mild dishwashing liquid and water
10% household bleach solution (1 part bleach in 9
parts water) made fresh daily
Super Sani-wipes (ammonium chloride with alcohol)
Do not get moisture into the code key slot, test strip slot,
or the serial port.
Note to Healthcare Professionals:
Follow the infection control procedures
appropriate for your facility.