User`s manual

4. Hold the control bottle
horizontally with the tip
pointed directly at the right
edge of the strip.
5. Gently squeeze the bottle to
form one small drop.
6.
Bring the drop to the edge of the strip within the curve,
and allow the strip to automatically draw the solution
into the yellow window until it is completely filled.
An hourglass flashes in the
display until the measurement
is complete.
7. Tightly replace the cap on
control solution.
8.
The control result appears. If result is within the
acceptable range (which is found on the side of the
vial of strips), an icon of the control vial appears on
the screen while the control test result and time and
date (if set) alternate with “OK”and “L1” or “L2”on
the display.
Example of an acceptable Level 1 control result.
Performing a Quality Control Check
1. Open test strip vial and remove test strip.
Replace vial cap.
2. Within 3 minutes,
insert
test strip (yellow
window
facing up) into test strip
slot.The meter turns on
automatically.
Check that the code
number on the meter
matches the code on the
vial of test strips you are
using.The test strip symbol
stops flashing and a blood
drop symbol appears on the display.
3. Press and release the right arrow button once to
select a L1 (Level 1) control or twice to select a
L2 (Level 2) control.
Note: Results marked as L1 or L2 do not appear
in memory or averages.