ACCU-CHE HEK Advantage BLOOD GLUCOSE METER User’s Manual ®
Welcome Thank you for selecting the Accu-Chek Advantage System. Designed for ease of use, this system was produced in close association with diabetes healthcare professionals, hospitals and, most importantly, people with diabetes. The Accu-Chek Advantage System is intended for self-testing of glucose in capillary whole blood by people with diabetes. Suitable for self-testing. Your new meter is designed for testing fresh capillary whole blood samples (for example, blood from your fingertip).
The Accu-Chek Softclix finger pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk. Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline -- Third Edition; CLSI document M29-A3, 2005).
Contents About the Accu-Chek Advantage Meter . . . vi Special Features . . . . . . . . . . . . . . . . . . . . viii About the Accu-Chek Advantage Plus Test Strip . . . . . . . . . . . . . . . . . . . . . . . . . . . ix Inserting the Battery . . . . . . . . . . . . . . . . . . 1 Battery Symbol on the Display . . . . . . . . . . 2 Setting the Beeper, Time, Date, and Hypo Flag . . . . . . . . . . . . . . . . . . . . . . . 4 Turning the Beeper On or Off . . . . . . . . . . . 4 Setting the Time and Date . . . .
Patient Testing for Healthcare Professionals . . . . . . . . . . . . . . . . . . . . . . . Sample Handling . . . . . . . . . . . . . . . . . . . Testing Blood Glucose in Multiple Patients . Performing a Blood Glucose Test . . . . . . . . Flagging Test Results . . . . . . . . . . . . . . . . Normal Blood Glucose Values . . . . . . . . . . Unusual Test Results. . . . . . . . . . . . . . . . . Troubleshooting Checks . . . . . . . . . . . . . . Action. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error Messages . . . . . . . . . . . . . . . . . . . . . Technical Data . . . . . . . . . . . . . . . . . . . . . . Product Specifications . . . . . . . . . . . . . . . Electromagnetic Compatibility . . . . . . . . . Performance Analysis . . . . . . . . . . . . . . . . Note on Patient Situation . . . . . . . . . . . . . Safety Remarks . . . . . . . . . . . . . . . . . . . . . About the Accu-Chek Softclix Finger Pricker . . . . . . . . . . . . . . . . . . . . . . Accu-Chek Softclix Finger Pricker . . . . . .
About the Accu-Chek Advantage Meter ON/OFF Button Press to turn meter ON or OFF. Serial Port Used to transfer data. Display Shows blood glucose result, messages, and glucose results stored in memory. Right and Left Buttons Used for meter setup and review of memory. Test Strip Slot Insert test strip here.
About the Accu-Chek Advantage Meter Code Chip For coding the meter. A new code chip comes with every container of test strips. Battery Cover Remove cover to change battery. Code Chip Slot Insert code chip here.
Special Features The meter is designed to store information important to you and your healthcare professional. The memory stores 480 results. Along with these results, the meter can store additional important information. The hypo flag allows you to define your hypoglycemic level and receive a reminder whenever you obtain a blood glucose result below this level. See “Setting the Hypo Flag” for additional information.
About the Accu-Chek Advantage Plus Test Strip Electrodes Insert test strip with silver-coloured bars up and toward the meter. Yellow Window Do not apply blood on the top of the test strip. Before Applying Blood – The window should be completely yellow. After Applying Blood – The window should be completely filled with the blood drop. Note: Any area of the test strip can be touched without affecting the glucose result. Comfort Curve Touch the blood drop to the edge of the test strip.
Inserting the Battery The meter is shipped with one 3-volt battery type 2032 that needs to be inserted before testing. The battery that comes with the meter can be found in the mesh pocket of the carrying case. Battery life varies depending on usage, so always keep a spare packaged battery on hand. The meter saves battery power by automatically turning off after 90 seconds of non-use or five seconds after taking a test strip out of the meter following a test.
2. Insert the 3-volt battery (type 2032) into the compartment with “+” side facing you. 3. Snap the battery cover back in place. If the time and date feature was set previously, the meter turns on automatically. Confirm that the time and date are set correctly. If the time and date had not been set previously, wait five seconds then turn the meter ON. The meter beeps if the beeper is turned on.
If a flashing battery symbol appears on the display, the battery is nearly dead. Battery must be replaced to continue testing. After changing the battery, the meter prompts you to verify the time and date immediately after the battery is inserted, if time and date had previously been set. See “Setting the Time and Date.” Note: To retain the correct time and date settings, replace the battery before it is completely dead. Please dispose of the meter battery according to your local ordinances.
Setting the Beeper, Time, Date, and Hypo Flag Turning the Beeper On or Off Does the beeper have to be turned on? No, it is an optional feature. Turning the beeper on or off does not affect the test results. If the beeper is turned off, beeps that occur due to errors still sound.
3. Press and hold down the ON/OFF button until “set” appears at the bottom of the screen. Release the ON/OFF button. “On,” “set,” and the flashing beeper symbol appear on the display. The beeper is now on. Note: You may leave the set mode at any time by pressing and holding the ON/OFF button. The settings you have entered are saved. 4. Press and release the right or left arrow button. “OFF,” “set,” and the beeper symbol appear on the display. The beeper is now off.
5. Press and hold down the ON/OFF button until the test strip symbol flashes or code and “---” appear to leave this function. Setting the Time and Date Do I have to set the time and date? Time and date should be set prior to testing if you plan to download the results into software or if you plan to review the memory to check when a test was done. To set the time and date: 1. Press and release the ON/OFF button to turn the meter ON. 2.
5. Press and release the ON/OFF button. The hour display flashes. 6. Press and release the right or left arrow button until the correct hour is displayed. 7. Press and release the ON/OFF button. The minute display flashes. 8. Press and release the right or left arrow button until the correct minute is displayed. 9. Press and release the ON/OFF button. The month display flashes. 10. Press and release the right or left arrow button until the correct month is displayed. 11. Press and release the ON/OFF button.
Setting the Hypo Flag The hypo flag allows you to define your hypoglycemic level and receive a reminder whenever you obtain a blood glucose result below this level. You can set the hypo indicator level to any number between 3,3 and 4,4 mmol/L. Important Note: Consult with your healthcare professional to set the hypo marker to compliment your therapy. Do not use the hypo marker as the only indicator of hypoglycemia. Rely upon your symptoms and directions from your healthcare professional.
5. Press and release the right or left arrow button to turn ON the hypo flag. 6. Briefly press and release the ON/OFF button. “Hypo” flashes, and the current hypo indicator setting, “set,” and “mmol/L” appear on the display. 7. Press and release the right or left arrow button to adjust the hypo flag setting. 8. To leave this function, press and hold down the ON/ OFF button until the flashing test strip symbol or code and “---” are displayed.
The code chip is properly inserted when it snaps into place. You need to insert a new code chip: • Whenever you open a new box of test strips. • Whenever either one of these displays appears: 1. Make sure the meter is turned off. Turn the meter over so you are looking at the back. Remove the old code chip if one is installed. 2. Insert the new code chip until it snaps into place. 3. Turn the meter ON. A 3-digit code number appears. This number must match the code number on your container of test strips.
Testing Gathering Supplies To begin testing, you need the following items: • Accu-Chek Advantage Meter • Accu-Chek Advantage Plus Test Strips • Accu-Chek Softclix Lancet • Accu-Chek Softclix Finger Pricker Performing a Test Healthcare professions please see “Testing Blood Glucose in Multiple Patients.” 1. Wash your hands in warm, soapy water. Rinse well and dry completely. Warming fingers can increase blood flow. 2. Load the finger pricker according to manufacturer’s instructions.
Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline -Third Edition; CLSI document M29-A3, 2005). 3. Remove a new test strip from the container. Be sure to tightly replace the container cap after removing the test strip. 4.
5. Let your arm hang down at your side briefly to allow blood to flow to your fingertips. 6. Grasp your finger just below the joint closest to the fingertip, and squeeze for three seconds. 7. Obtain a blood sample. Use the lancet to obtain a drop on the side of your finger. Do not obtain the drop from the pad of your finger. To perform the test, obtain a small drop of blood resting on your finger. 8. Touch and hold drop of blood to the edge of the test strip within the curve.
9. When blood is applied to the test strip, an hourglass symbol flashes on the display until the measurement is completed. The result appears on the display. 10. The blood glucose result is displayed in mmol/L. If your meter displays mg/dL contact the Accu-Chek Customer Careline on 0800 701 000 (UK) or 1 800 709 600 (Ireland). Using the wrong meter can cause you to wrongly interpret the test result. The result is automatically recorded in the meter’s memory.
11. Remove the test strip from the meter. Discard the used test strip and lancet in a punctureproof container with a lid. If you have completed your testing, turn the meter OFF, or the meter automatically turns off five seconds after removing the test strip. If HI is displayed, your blood glucose result may be higher than 33,3 mmol/ L. (Refer to “Understanding Your Test Results.”) If LO is displayed, your blood glucose result may be lower than 0,6 mmol/L. (Refer to “Understanding Your Test Results.
Marking a Result with a Flag, Asterisk ( ) This feature allows you to perform a test and mark the result with a flag (asterisk) to be stored in memory. Each result flagged with the asterisk has special meaning only to you. For your healthcare professional, you should record the meaning along with your reading in your Record Book or Diary. An example includes using the asterisk to mark a result taken immediately after a meal. To mark a result with a flag (asterisk): 1.
Normal Blood Glucose Values These test strips are calibrated to deliver whole blood results. The normal fasting blood glucose range for an adult without diabetes is 3.5-5.3 mmol/L.1 For people with diabetes: Please consult your healthcare professional for the blood glucose range appropriate for you. You should treat your low or high blood glucose as recommended by your healthcare professional.
Then perform a quality control check with glucose control solutions and a new test strip. (See “Quality Control Checks.”) If the control result is within the acceptable range, review proper testing procedure and repeat your blood glucose test with a new test strip. If your blood glucose value is still inconsistent with your previous results, glucose trends, or how you feel, please contact your healthcare professional. Follow the advice of your healthcare professional before you change your therapy.
Correct Test Strip Appearance Although only a small amount of blood is needed, it is very important that you put enough blood on the test strip so that the entire yellow window is filled with blood. This ensures that the meter can produce accurate and reliable results. This test strip has the correct amount of blood. Notice that the left side of the yellow window is not completely filled with blood. This test strip has too little blood and may produce an error message (or an inaccurate result).
Using the Memory The meter automatically stores 480 glucose results, letting you review them in order from the most recent to the oldest. If you have set the time/date feature, the time and date of the results are also displayed. If time and date are not set, each result is marked with a number according to how it falls in sequence. If the memory is full and a new result is added, the meter deletes the oldest result. Results stored in memory are not lost when you replace the battery.
chip is installed, the code number appears briefly, followed by the most recent result stored in memory. If no code chip is installed, code and “---” appear. Press and release the right or left arrow button again to enter memory recall mode. 2. Once in the memory mode: Press the left arrow button to view individual results in memory, Or Press the right arrow button to review, in sequence 7-, 14- and 30-day averages.
Quality Control Checks What are quality control checks? They ensure that you are doing a test correctly and that your system is working properly. When should I check my meter? • If you leave the cap off the container of test strips. • Before using the meter for the first time. • When you open a new container of test strips. • If you drop the meter. • Whenever your result does not agree with the way you feel. • If you have repeated a test, and the blood glucose result is still lower or higher than expected.
To do a quality control check, you need: • Accu-Chek Advantage II Glucose Control Solutions (Level 1 and Level 2) • Accu-Chek Advantage Plus Test Strips Note: When you open a new bottle of glucose control solution, write the date on the label. Opened bottles of glucose control solution are good for three months or until the expiration date, whichever comes first.
Performing a Quality Control Check 1. Open the test strip container and remove a test strip. Replace the container cap. 2. Within three minutes, insert the test strip (yellow window facing up) into test strip slot. The meter turns on automatically. Check that the code number on the meter matches the code on the container of test strips you are using. The test strip symbol stops flashing and a blood drop symbol appears on the display. Cod e: 3.
5. Gently squeeze the bottle to form one small drop. 6. Bring the drop to the edge of the test strip within the curve, and allow the test strip to automatically draw the solution into the yellow window until it is completely filled. An hourglass symbol flashes in the display until the measurement is complete. 7. Tightly replace the cap on control solution. 8. The control result appears.
Note: Result must be marked with an L1 or L2 for the meter to verify if the control result is within range. If result has not been marked with an L1 or L2, refer to the test strip container label for the glucose control ranges. If result is not within the acceptable range, a control bottle symbol appears on the screen while the control test result, and time and date (if set), alternate with “Err” and “L1” or “L2” on the display. See “Understanding Control Results” for more information.
Understanding Control Results If your glucose control results are within the acceptable range, you can feel confident that the test strips and meter are working correctly. If one or both of your glucose control results are outside of the acceptable range, the system may not be working properly.
Note: Consumers – please proceed to the “Caring for Your Meter” section. The following sections are for healthcare professionals only. Patient Testing for Healthcare Professionals Sample handling Always wear gloves and follow your facility’s infection control procedures when handling blood-contaminated items. Always adhere to the recognized procedures for handling objects that are potentially contaminated with human material. Follow the health and safety policy of your laboratory or institution.
Testing Blood Glucose in Multiple Patients Only healthcare professionals are allowed to perform blood glucose tests in a multiple patient testing situation. When using the meter always follow the recognized procedures for handling objects that are potentially contaminated with human material. Practise the health and safety policy of your laboratory or institution. Also, to stop your meter becoming contaminated, please follow the directions printed on the next few pages.
The Accu-Chek Softclix finger pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk. Any object coming into contact with human blood is a potential source of infection (see: Clinical and Laboratory Standards Institute: Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition; CLSI document M29-A3, 2005).
4. Use a finger pricker with adjustable penetration depth settings. If using a Safe-T-Pro Plus finger pricker, select a penetration depth to suit the nature of the patient’s skin. Start with a shallow depth setting. 5. Select a puncture site along the side of a fingertip. 6. Insert a test strip into the meter. 7. Make sure the code number on the meter display matches the code number on the test strip container. 8. Lance the side of the fingertip. 9.
Flagging Test Results It is possible to flag a test result with an asterisk ( ) for a special result. When the results in memory are reviewed, this flag is a reminder of a special result. 1. Perform a test. 2. With the test result on the display, and the test strip STILL IN THE METER, press the right or left arrow button once. The appears in the lower right corner. 3. Remove and discard the used test strip and blood collection device per facility infection control policy.
Dispose of used lancets, test strips or disposable finger prickers according to local health and safety regulations. Note: The test strip cannot properly draw up blood that has spread over the skin. Normal Blood Glucose Values These test strips are calibrated to deliver plasma-like results. The normal fasting blood glucose range for an adult without diabetes as related to plasma is 4.1-5.9 mmol/L.1 Blood glucose concentrations may be measured in whole blood or in plasma.
Troubleshooting Checks 1. Check if the test strips were expired. 2. Check if the cap on the test strip container was always closed tightly. 3. Check if the test strip was out of the container for a long time. 4. Check if the test strips were stored in a cool, dry place. 5. Check if the testing steps were followed. 6. Check if the code number on the meter display matches the code number on the test strip container. 7.
4. Repeat the test with a properly stored test strip. 5. Read “Patient Testing for Healthcare Providers” and test again. If you still have problems, call Roche. 6. If they do not match, insert the correct code chip into the meter and test again. 7. Repeat the test with a new test strip and perform a control test. If the problem does not resolve, call Roche. Disinfecting the Meter The Accu-Chek Advantage meter may be disinfected.
How to Disinfect • Make sure the meter is off. • Gently wipe the meter’s surface. • Cotton buds, wipes and cloths lightly moistened with disinfectant are all suitable for disinfecting the meter. • Allow areas you have wiped time to dry thoroughly. • Remove any fluff or lint that may remain. • Refer to the section “Cleaning the Meter” for further information. Notes: Make sure that no liquid enters the meter. Do not spray anything onto the meter and do not immerse it in liquid.
If blood has entered the meter, wipe the area around the test strip slot and the display screen. Allow areas you have wiped time to dry thoroughly. The meter, including the test strip slot and display screen, must be thoroughly cleaned and disinfected at least once a day after use. Carefully check all recesses, grooves and gaps.
Control Testing for Healthcare Professionals Introduction Testing control solutions with known glucose levels establishes that the operator and the system are performing acceptably. Control results must be within the defined acceptable ranges before valid patient testing is allowed. Testing Intervals Follow your facility’s policy for control testing intervals.
Control results must be within the designated range on the test strip container label, or as defined by your facility, before being considered acceptable. Patients can be tested after controls have been acceptably performed at the proper testing interval. About the Control Solutions • Only use Accu-Chek Advantage Control Solutions. • Write the date the bottle was opened on the bottle label.
3. Press and release the right arrow button once to select a L1 (Level 1) control or twice to select a L2 (Level 2) control. Note: Results marked as L1 or L2 do not appear in memory or averages. 4. Hold the control bottle horizontally with the tip pointed directly at the right edge of the test strip. 5. Gently squeeze the bottle to form one small drop. 6.
If result has not been marked with an L1 or L2, refer to the test strip container label for the glucose control ranges. and Example of an acceptable Level 1 control result. If result is not within the acceptable range, a control bottle symbol appears on the screen while the control test result, and time and date (if set), alternate with “Err” and “L1” or “L2” on the display. See “Understanding Control Results” for more information. and Example of an unacceptable Level 1 control result. 9.
Caring For Your Meter Display Check This verifies that all parts of the display are working. Make sure the meter is off. Press and hold down the ON/OFF button. The display should look exactly like the picture at right. If it does not, call Roche Diagnostics.
Storage and Handling Keep the meter free of dust. Protect it from extremes in temperature and humidity, and handle it with care. If the meter is dropped, perform a quality control check with the glucose control solutions before doing a blood glucose test. The travel case is designed to let you store a variety of supplies you may need and helps to protect the meter.
Quick Reference Display Guide Meter needs to be coded or code chip is not inserted completely; turn meter OFF and insert code chip. Meter is in the “set” function, allowing you to set the beeper, time, date, year, and hypo flag. Meter is ready to accept a test strip. Blood drop can be applied to test strip.
Blood glucose result appears on display. Blood glucose result may be higher than 33,3 mmol/L. If this does not confirm the way you feel, review proper testing procedure and perform a quality control check. Repeat test. If HI still appears, follow your healthcare professional’s instructions. Blood glucose result may be lower than 0,6 mmol/L. If this does not confirm the way you feel, review proper testing procedure and perform a quality control check. Repeat test.
Temperature is above or below the operating range of test strips (14 ° to 40 °C). The result you have obtained may not be accurate. Move to an area between 14 ° and 40 °C, wait five minutes, and repeat the test. Do not artificially heat or cool the meter. Memory Displays The meter is in the “memory” function, allowing you to recall results from memory.
If you are reviewing memory and a result appears with a temperature symbol as shown, this result may not be accurate because the test was performed outside the operating range of the test strips (14 ° to 40 °C). The meter is in the “memory” function, allowing you to recall 7-, 14-, and 30-day averages. This blood glucose result was flagged (asterisk) . Do not use individual values stored in memory to make therapy adjustments.
Display Messages Battery Displays Battery power is low. Change battery soon. (battery symbol not flashing) Battery is dead. Change battery now. Meter turns off after three seconds.
Quality Control Check Displays Meter is ready for a Level 1 or Level 2 glucose control check. Apply a drop of Level 1 or Level 2 glucose control solution.
and Example of an acceptable control result. Glucose control result is not acceptable. Repeat test. Refer to Understanding Control Results (see page 27). and Example of an unacceptable control result.
Error Messages The meter automatically performs a system check to prevent inaccurate results. If the meter detects a malfunction, it displays the “Err” message and an appropriate symbol. Do not make treatment decisions based on an error message. The code chip is faulty, or the code chip was removed while the meter was on. Turn meter OFF and re-insert code chip. If this error display still appears, contact Roche Diagnostics.
Your blood glucose may be extremely high or a meter or test strip error has occurred. If this confirms how you feel, contact your healthcare professional immediately. If it does not confirm how you feel, repeat the test and refer to the section in the manual titled Understanding Your Test Results. If this does not confirm the way you feel, perform a quality control check with your glucose control solutions and a new test strip.
Meter has detected the loss of all results in memory. Turn meter OFF and turn it back ON again. You may not be able to recall the results previously stored in memory, although your meter will store any future test results. I I I II I I I I I I Temperature is above or below the operating range of the meter. Move to an area between 14 ° and 40 °C, wait five minutes, and perform a test. Do not artificially heat or cool the meter. Meter has failed its series of electronic checks.
Technical Data Please read information packaged with test strips regarding up to date product specifications and limitations. Measuring Principle: The glucose dehydrogenase converts the glucose in a blood sample to gluconolactone. This reaction liberates an electron that reacts with a coenzyme electron acceptor, the oxidized form of the mediator hexacyanoferrate (III), forming the reduced form of the mediator, hexacyanoferrate (II). The test strip employs the electromechanical principle of biamperometry.
Product Specifications Blood Volume 4μL Sample Type Fresh capillary whole blood Measuring Range System measuring range is 0.6–33.3 mmol/L. Test Strip Storage Conditions Store test strips at room temperature between 2 ° to 32 °C. Do not freeze. Accuracy Refer to your Test Strip Package Insert for more information. Repeatability Refer to your Test Strip Package Insert for more information. Meter Storage Conditions -25 ° to 70 °C <85% humidity >53.
Construction Hand held Protection Class III Rated Operation Continuous operation Data Port 3-wire RS-232C Serial Port Electromagnetic Compatibility This meter meets the electromagnetic immunity requirements as per ISO 15197 Annex A. The chosen basis for electrostatic discharge immunity testing was basic standard IEC 61000-4-2. In addition, it meets the electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
The Accu-Chek Advantage System complies with the requirements of EN ISO 15197. References 1. Stedman, TL. Stedman’s Medical Dictionary, 27th Edition, 1999, p. 2082. Note on Patent Situation The Accu-Chek Softclix Finger Pricker is covered by European Patent Nos. 1 034 740, 0 565 970, and 0 782 838. Safety Remarks Only use an Accu-Chek Softclix Lancet with the Accu-Chek Softclix Finger Pricker since the use of other lancets may prevent or impair proper function of Accu-Chek Softclix Finger Pricker.
A B C D E G F H I J 58
About the Accu-Chek Softclix Finger Pricker A Cap B Depth selection arrow C Notch D Protective cap E Accu-Chek Softclix lancet F Lancet holder G Semi-circular cutout H Ejector I Transparent release button J Priming button 59
Accu-Chek Softclix Finger Pricker Introduction The Accu-Chek Softclix finger pricker is an easy-to-use finger pricking device for obtaining blood from the tip of the finger or the earlobe virtually without pain. Penetration depth can be adjusted precisely to suit individual skin requirements. The Accu-Chek Softclix finger pricker is intended for patient self-monitoring by a single person only. It must not be used to collect blood from more than one person as this poses an infection risk.
For this purpose the Accu-Chek Softclix finger pricker can be set to different penetration depths ranging from 0.5 to 5.5. They are indicated on the cap by the numbers 1 to 5. The additional intermediate settings marked with dots enable a particularly precise adjustment of Accu-Chek Softclix finger pricker to suit individual skin type. The markings on the device show the depth of penetration selected. Penetration depth can be altered by turning the front part of the cap.
1 2 3 4 5 6 7 62
How to Take a Blood Sample 1. Pull off the cap from the Accu-Chek Softclix finger pricker. 2. Insert a new Accu-Chek Softclix lancet as far as it goes into the lancet holder and push it in until it clicks. 3. To remove the protective cap, twist and then pull it. 4. Press the Accu-Chek Softclix finger pricker cap on again until it clicks. Be sure to align the notch on the cap with centre of the semicircular cut-out on the ejector. 5. First check that you have set the desired penetration depth.
Wait a few seconds to allow the puncture site to open so that a drop of blood can form. Apply the blood drop to the test strip. Then carry out the test according to the instructions that come with the test strips or meter. 7. To remove the lancet, pull off the Accu-Chek Softclix finger pricker cap. To eject the used lancet slide the ejector forward to remove it, and dispose of the lancet in a sealed container. To avoid infections use a new lancet each time you take a blood sample.
Subsequent Treatment of the Puncture Site Wipe the puncture site with a dry, clean tissue. If it continues to bleed, apply gentle pressure to the puncture site with a plaster to protect it from dirt or infection. Cleaning and Disinfecting the Accu-Chek Softclix Finger Pricker To clean the Accu-Chek Softclix finger pricker, use any mild liquid detergent and a soft cloth. Avoid immersing the device completely in liquid.
Product Safety Information Strong electromagnetic fields may interfere with the proper operation of the meter. Do not use this meter close to sources of strong electromagnetic radiation. To avoid electrostatic discharge, do not use the meter in a very dry environment, especially one in which synthetic materials are present. Disposing of the Meter During blood glucose measurement, the meter itself may come into contact with blood. Used meters therefore carry a risk of infection.
Please consult instructions for use Caution (refer to accompanying documents). Please refer to safety-related notes in the manual accompanying this instrument. Store at Use by / Expiry date unopened/ opened Discard three months after opening or by the expiry date, whichever comes first. Manufactured by Catalogue number Lot number Listed by Underwriter’s Laboratories, Inc.© in accordance with UL 61010A-1 and CAN/CSA C22.2 No.1010-1. For in vitro diagnostic use.
Dispose of in domestic waste 3-volt lithium battery (coin cell 2032) Accu-Chek Softclix Finger Pricker On the finger pricker itself and on the packaging of Accu-Chek Softclix Lancet you will encounter the following symbols, shown here with their meaning. Please read this note with special care For single use only The lancets may only be used up to the expiry date specified on the box.
Warranty and Service Instructions Roche Diagnostics warrants the meter against any material and manufacturing defects that may arise during proper use of the instrument. For precise terms and conditions please see the Warranty Card in the box. This warranty does not affect any statutory or other rights. The warranty ceases to apply if the instrument has been misused or tampered with.
Accu-Chek Softclix Finger Pricker Guarantee The Accu-Chek Softclix Finger Pricker is guaranteed for life from the date of purchase against defects in materials or workmanship. Please fill in the enclosed Warranty Card and return it to Roche Diagnostics Limited. Our liability is explicitly limited to the elimination of demonstrable manufacturing defects. This guarantee is rendered invalid if the instrument has been improperly operated or otherwise tampered with.
Index About the Accu-Chek Softclix Finger Pricker ..........59 About the Accu-Chek Advantage Meter ........................ vi Accu-Chek Advantage Plus Test Strip ..................... ix, 11 Accu-Chek Softclix Finger Pricker Introduction ............60 Battery ..................................................................1, 2, 3 Battery display .......................................................... 2, 3 Battery life.....................................................................
Error messages ..........................................................51 Flag (asterisk) ........................................................... 16 Gathering supplies for testing ................................. 11 Getting a good sample .................................................11 Glucose control solutions ............................................ 23 HI display ..............................................................15, 31 How to take a blood sample...................................
Testing, blood glucose ............................................. 11 Testing, quality control ......................................... 22, 49 Test strip slot ............................................................... vi Test strip container .............................................. 10, 12 Time and date, setting .................................................. 6 Transferring data..........................................................43 Troubleshooting ......................................
Any Questions We want to give you the help you need for long and reliable service of your meter. Review your test procedure and make sure that you are performing the steps correctly. If you still have questions, please contact Roche Diagnostics. Please have your Accu-Chek Advantage Meter, Accu-Chek Advantage Plus Test Strips and Accu-Chek Advantage II Glucose Control Solutions available when you call. IMPORTANT—READ CAREFULLY BEFORE USE.
75
76
This product fulfills the requirements of Directive 98/79/EC on in vitro diagnostic medical devices.
ACCU-CHEK, ADVANTAGE, and SOFTCLIX are trademarks of Roche. Accu-Chek Customer Careline* 0800 701 000 (UK) 1 800 709 600 (Ireland) *Calls may be recorded for training purposes Distributed by: Roche Diagnostics Limited Charles Avenue Burgess Hill West Sussex RH15 9RY 00 + 44 (0) 1444 256000 Roche Diagnostics GmbH, 68298 Mannheim, Germany www.accu-chek.