Prüfmittel-Nr. 43 IX ACCU-CHEK Active BLOOD GLUCOSE MONITOR User’s Manual ACCU-CHEK and SOFTCLIX are trademarks of a Member of the Roche Group. Roche Diagnostics GmbH D-68298 Mannheim, Germany www.accu-chek.com 015134 AccuChek Active_Eng.indd u2-u3 0 4493206001(01) 06/04 0088 12.08.
Contents Please read this manual carefully and completely before performing the first measurement. Should you have any questions, please contact one of our customer support and service centres. You will find a list of addresses in Chapter 9.4. On packaging, on the type plate and in the User’s Manual you may encounter the following symbols, shown here with their meaning: 1 For in vitro diagnostic use.
Contents 4 5 6 | || | | | | | | | || || | | | | | | || || | | Contents Measuring blood glucose with your Accu-Chek Active monitor 36 4.1 Materials for testing blood glucose 4.2 Preparing for the blood glucose test 4.3 Testing blood glucose In-monitor dosing Plausible results Implausible results Out-of-monitor dosing Testing by healthcare professionals 36 37 38 44 49 52 55 59 Using Accu-Chek Active as an electronic diary 60 5.1 Results memory 5.
Parts summary Accu-Chek Active Parts summary Accu-Chek Active Liquid crystal display (LCD) 1 M(emory) button 2 S(et) button 3 Test strip slot with switch 4 Measurement optics 5 Test strip guide 6 Battery compartment 7 Battery 8 Coding chip 9 8 7 1 3 2 9 5 4 6 4 015134 AccuChek Active_Eng.indd 4-5 5 12.08.
Parts summary Accu-Chek Active Parts summary Accu-Chek Active Coding chip slot 10 Infrared interface 11 10 11 6 015134 AccuChek Active_Eng.indd 6-7 7 12.08.
1 Introduction 1.1 The monitor’s main features at a glance Your Accu-Chek Active blood glucose monitor is a state-of-the art precision instrument that will enable you to test your blood glucose simply and with confidence.
1.1 The monitor’s main features at a glance 씰 Integrated data evaluation: From the results stored in memory your monitor can calculate the average of your blood glucose results over the last 7 or 14 days. 씰 Out-of-monitor dosing: After activating the test cycle you may remove the test strip to apply blood (e.g. from an earlobe), and then insert it again in the monitor.
1.2 About this user’s manual This User’s Manual will help you get to know your new AccuChek Active monitor step by step. It will provide you with all the information you need to operate, troubleshoot, clean and care for your monitor. It is important to remember that, in order to keep your monitor in tip-top condition, you need to comply with the care instructions in addition to following the correct operating procedures. Accu-Chek Active is a precision instrument.
2 Modes As you will have gathered, your Accu-Chek Active monitor can do much more than just measure blood glucose. It offers a number of additional functions, which fall into three groups. “Mode” refers to a group of functions that you can call upon. You select a mode simply by turning on the monitor or by the way in which you turn it on. Detailed information on the various ways of turning on your monitor and the options available in each mode is provided later in this manual.
2.2 Memory Mode 2.3 Set Mode Memory Mode provides you with various functions to help you evaluate your results. To enter this mode, turn on your Accu-Chek Active monitor by pressing the M(emory) button. To enter Set Mode, turn on your Accu-Chek Active monitor by pressing the S(et) button. 씰 You can retrieve up to 200 results. 씰 You can choose the format in which you wish the date and time to be displayed.
3 Initial steps before testing 3.2 Basic settings 3.1 After opening the pack Accu-Chek Active is sold complete with one lithium battery already inserted. A plastic film protects the battery against becoming drained prematurely. Before using the monitor, remove the film covering the battery. When you turn on your Accu-Chek Active monitor for the first time, it has the following settings (which you can change): 씰 Display format for the date: day.month/DD.MM (or alternatively month.
3.2 Basic settings Accu-Chek Active comes in 2 versions. When you first turn on your Accu-Chek Active, please be sure to check that your usual unit (mg/dL or mmol/L) is displayed above the large numerals in the display (888). Ask your doctor if you do not know which is the right unit for you. This setting cannot be changed. If you have a monitor showing the wrong unit, please ask your dealer for a replacement. 20 015134 AccuChek Active_Eng.
3.2 Basic settings Turning the monitor on in Set Mode You use only the M and S buttons on the front of your Accu-Chek Active to enter the desired settings. 씰 Briefly press the S button (for less than 3 seconds). Accu-Chek Active is now turned on and in Set Mode. In the display you see the following: 씰 the current time display format (24h/12h) 씰 flashing: the current time, or 0:00 (a.m.) when you turn on the monitor for the first time 씰 flashing: the current date, or 0.
3.2 Basic settings The way in which the date and time are displayed from here on depend on the chosen display format. 씰 If you wish to change the time display format, press the M button now. This allows you to choose between 24h and 12h display format. 씰 When you see the desired format, press the S button once to proceed to the date and time settings. 24 015134 AccuChek Active_Eng.indd 24-25 25 12.08.
3.2 Basic settings Setting the date and time The procedure for setting the date and time is essentially the same for all units of time (day, hour, etc.): 씰 Use the S button to select the unit of time you wish to change. The unit will flash, indicating that it is ready to be changed. 씰 Use the M button to modify the unit of time selected. 씰 Pressing the M button once will increment the value of the unit by one (one hour, one minute, etc.).
3.2 Basic settings 씰 Confirm the setting by pressing the S button again. The display moves on automatically to the next unit of time, which again flashes. 씰 Use the same procedure to set all the remaining units of time. The date and time are set in the following sequence: hours –> minutes –> year –> month –> day The year is already set. Press the S button to confirm. The monitor automatically recognises, and allows for, leap years. 28 015134 AccuChek Active_Eng.indd 28-29 29 12.08.
3.2 Basic settings Setting the beep tone After you have set the day and pressed the S button, you can either enable or disable the beep tone. 씰 Press the M button to turn the beep tone on and off. 씰 Press the S button to complete the settings procedure. Check that all the settings now shown on the display are correct. Press the S button again to turn off your Accu-Chek Active monitor. If any of the settings are incorrect, you may repeat the settings procedure (after turning the monitor on again).
3.3 Coding The properties of the Accu-Chek Active test strips you use with the monitor may differ very slightly from one pack to another (because of lot-to-lot differences). Your Accu-Chek Active monitor must be programmed to allow for these differences in properties in a process known as “coding”. All the information relating to the strips in a particular pack of Accu-Chek Active test strips is held in the coding chip that comes with that pack.
3.3 Coding Inserting the coding chip When coding, leave your Accu-Chek Active monitor turned off. 씰 Take the new coding chip from the pack of test strips. 씰 Gently slide the coding chip into the slot on the side of the monitor ➉ (see parts summary). 씰 Ensure that you push the coding chip until it clicks into place.
4 Measuring blood glucose with your Accu-Chek Active monitor 4.2 Preparing for the blood glucose test 4.1 Materials for testing blood glucose Please have the following items ready to carry out the test: 씰 your Accu-Chek Active monitor with the coding chip inserted, 씰 the pack of Accu-Chek Active test strips you took the coding If you are not yet familiar with the concept of blood glucose selfmonitoring using a monitor, practise first by carrying out a quality control test as described in Chapter 6.
4.3 Testing blood glucose Check the round control window on the back of the test strip with the colour scale printed on the test strip container. The colour of the control window must match that of the colour interval at the top (0 mg/dL / 0 mmol/L). If the test strip shows a different colour, do not use it. Hold the test strip so that the application area and arrows are facing upwards.
4.3 Testing blood glucose 씰 The monitor carries out a display test (for details see page 87), then shows the code number. Is this the number printed on the test strip container? If not, check that you really did insert the coding chip from the new pack. If you are using the right coding chip but the number in your display is wrong, please call your customer support and service centre. If “code” is flashing and you see three horizontal bars (---) instead of a number, you have not inserted the coding chip.
4.3 Testing blood glucose 씰 Check that the correct date and time are displayed. Do not carry out any tests if the code number in the display is not the same as the number on the test strip pack. Incorrect coding produces incorrect results. Sometimes the error can be significant, causing the wrong therapeutic decision to be taken and so producing adverse health effects. When the display test has been successfully completed and the code number matches, your Accu-Chek Active monitor is ready for testing.
4.3 Testing blood glucose In-monitor dosing If you have opted for in-monitor dosing, proceed as follows: 씰 Massage a fingertip to stimulate the circulation and make blood collection easier. 씰 Apply the drop of blood on to the middle of the orange-coloured, square application area (pad). It is OK to touch this area when applying blood, but do not attempt to spread the blood. If you suspect the amount of blood may not suffice, you may apply a second drop within 5 seconds.
4.3 Testing blood glucose 씰 The hourglass symbol indicates that testing is in progress. 씰 The monitor beeps again after about 5 seconds to signal that the test is complete. The result appears in the display and is automatically saved together with the date and time. 씰 Remove the test strip from the monitor, which will now auto- matically turn off. 46 015134 AccuChek Active_Eng.indd 46-47 47 12.08.
4.3 Testing blood glucose Plausible results Your Accu-Chek Active monitor measures blood glucose within a defined range (10-600 mg/dL or 0.6-33.3 mmol/L). Values outside this range are displayed as “Lo”- meaning the result is less than 10 mg/dL or 0.6 mmol/L - or “Hi”- meaning the result is greater than 600 mg/dL or 33.3 mmol/L. If you applied too little blood, please repeat the test with a new test strip. Now perform a plausibility check.
4.3 Testing blood glucose 씰 Turn over the test strip to reveal the round control window on the back. 씰 On the label of the test strip container is a colour scale with blood glucose values printed alongside. Select the blood glucose value that best approximates the reading you obtained. 씰 Compare the colour of the control window with the colour you selected on the label. The colours must be a fairly close match. If there is a great disparity, repeat the test.
4.3 Testing blood glucose Implausible results 씰 Was the test strip container you are presently using kept properly closed? If the displayed result does not reflect the way you feel or seems unusually high or low, or if colours do not match, please check the following: 씰 Have the test strips expired? 씰 Have the test strips been kept under suitable temperature 씰 With your Accu-Chek Active monitor, are you using only conditions (i.e.
4.3 Testing blood glucose Out-of-monitor dosing The test procedure is mostly identical with “Testing blood glucose” on page 38. Follow the instructions given there up to the point where you are instructed to apply blood. 씰 Clean and massage the site you intend to collect blood from (for example, an ear lobe). This will stimulate the circulation and make blood collection easier. 씰 Use the lancing device to prick the selected site. 씰 Without pressing too hard, encourage a drop of blood to form.
4.3 Testing blood glucose 씰 Pull the strip out of the monitor. The two symbols (blood drop and test strip) are flashing. You now have 20 seconds to apply blood to the test strip and re-insert it before the monitor switches off. After 15 seconds a beep tone once per second indicates that the time for measurement has expired. 56 015134 AccuChek Active_Eng.indd 56-57 씰 Apply the drop of blood to the middle of the orange-coloured pad.
4.3 Testing blood glucose 씰 Gently push the test strip back into the test strip slot of your Accu-Chek Active monitor, until you feel it click into place. 씰 Your Accu-Chek Active monitor beeps briefly to acknowledge application of blood and to announce the start of testing. Make sure you do not move the test strip during testing, or an incorrect result will be obtained. 씰 The hour glass symbol indicates that testing is in progress.
5 Using Accu-Chek Active as an electronic diary 5.1 Results memory Your Accu-Chek Active monitor has a 200-value memory for saving results together with date and time. You do not need to save a result specially; the monitor does it all for you. Once all the memory locations are full, the monitor automatically erases the oldest entry. All results in memory are consecutively numbered from 1 (most recent) to a maximum of 200 (oldest). 60 015134 AccuChek Active_Eng.
5.2 Retrieving results To use the memory functions of your Accu-Chek Active monitor, you must turn it on in Memory Mode. 씰 Briefly press the M button (for less than 3 seconds). In the display you see the following: 씰 at top left, “mem” (short for “memory”) 씰 followed by the most recent result (in memory location “1”), together with the time and date. 62 015134 AccuChek Active_Eng.indd 62-63 63 12.08.
5.2 Retrieving results To recall earlier results (in memory locations 2-200): 씰 Press the S button once to move back one result in time. As long as you keep the button pressed, the number of the memory location is displayed. As soon as you release the button, the result appears. 씰 By holding down the S button, you can scroll quickly through the results. As soon as you release the button, the result saved in that location appears.
5.2 Retrieving results Finding the 7 or 14 day average of glucose results As well as just showing the results in memory, your Accu-Chek Active monitor can calculate and display average values for the last 7 or 14 days. 씰 Start by pressing the M button to turn on your Accu-Chek Active monitor in Memory Mode. 씰 Then press the M and S buttons at the same time. The display shows “7 day ave” at top right. The figure shown is the average of glucose results for the last 7 days. 66 015134 AccuChek Active_Eng.
5.2 Retrieving results 씰 Then press the M and S buttons again at the same time. 씰 The display changes; the figure shown is now the average value for the last 14 days. To exit from the “average” screen, press the S button. The display reverts to the last result shown before the average value. Averages can only be calculated as long as you have set the date and time (see Chapter 3.2). Without these settings the monitor cannot identify the results from the last 7 or 14 days.
5.2 Retrieving results Erasing results from memory You can clear the current value (and only this) from memory. This feature is useful, for instance, if you have obtained an implausible result. Both the result and the time and date information are erased, though the memory location remains occupied. An average value cannot be displayed when the most recent result has been erased from memory. Results arising after a value has been erased are included in calculation of the average.
5.2 Retrieving results longer than 3 seconds 씰 Keep the M and S buttons pressed together for longer than 3 seconds. 씰 The current result is displayed. longer than 5 seconds 씰 Keep the M and S buttons pressed together for longer than 5 seconds. You will hear 5 beeps. 씰 All information relating to the current result is cleared and set to zero. The memory location remains occupied. Your Accu-Chek Active monitor turns off automatically after you have cleared the result. 72 015134 AccuChek Active_Eng.
5.3 Downloading data to a PC For data downloading you need a suitable PC program as well as the Accu-Chek Infrared Cable. For further information please contact your local Roche Diagnostics representative or your dealer. For further information please contact your local Roche Diagnostics representative or your dealer. Data can only be evaluated on a PC as long as you have set the full date and time (see Chapter 3.2).
5.3 Downloading data to a PC Downloading procedure To download data, proceed as follows: 씰 Set up the program and infrared cable according to manufacturer’s instructions so you can start data transfer with a mouse click. 씰 Position your Accu-Chek Active monitor about 10 cm from the infrared cable. Point the two infrared windows towards one another.
5.3 Downloading data to a PC 씰 Start downloading at the PC. The PC sends an instruction to the Accu-Chek Active monitor. 씰 During active downloading, “PC” remains constant in the display 씰 On completion of downloading, “End” appears in the display. Following the download you can turn off your Accu-Chek Active monitor with the M button, if the PC has not already shut down the monitor. In the latter case the “End” message may not be displayed. 78 015134 AccuChek Active_Eng.indd 78-79 79 12.08.
6 Checking Accu-Chek Active 6.1 Quality control testing with control solution To ensure your results are always dependable, you must periodically check the performance of your Accu-Chek Active monitor. Perform a quality control test: 씰 after you have opened a new pack of Accu-Chek Active test strips and consequently replaced the coding chip, 씰 after you have inserted a new battery, 씰 after you have cleaned your Accu-Chek Active monitor, 씰 if you have reason to doubt the validity of a result.
6.2 Preparing for the quality control test For the quality control test, please have the following items ready: 씰 your Accu-Chek Active monitor with the coding chip inserted, 씰 the pack of Accu-Chek Active test strips you took the coding chip from, 씰 the Accu-Chek Active Control solutions. Carefully read the pack inserts that came with the test strips and the control solutions, and select a control solution. 82 015134 AccuChek Active_Eng.indd 82-83 83 12.08.
6.3 Quality control test procedure 씰 Remove a test strip from its container. Close the container immediately. The cap contains a drying agent which ceases to function if the container is left open, rendering the test strips unusable. 씰 Check the round control window on the back of the test strip against the colour scale printed on the test strip container. The colour of the control window must match that of the colour interval at the top (0 mg/dL / 0 mmol/L).
6.3 Quality control test procedure 씰 The monitor performs a display test lasting 2 seconds. Check that all the segments making up the numerals (“888” or “88.8”) are properly displayed. If a segment is missing, test results may be displayed inaccurately (e.g. through 9 being confused with 3). If this happens, call your customer support and service centre. 씰 The current code number then appears in the display.
6.3 Quality control test procedure When the display test has been successfully completed and the code number matches, your Accu-Chek Active monitor is ready for testing. The display screen that follows signals: 씰 That the test strip has been inserted. 씰 The flashing drop symbol is your cue to apply the control solution (blood in the case of a real test). To make quality control results stand out later from blood glucose test results, you can place what is known as a “flag” against them.
6.3 Quality control test procedure 씰 Apply one drop of the selected control solution to the test pad of the strip. 씰 Your Accu-Chek Active monitor beeps briefly to acknowledge application of solution and to announce the start of testing. 씰 After 5 seconds a second beep tone indicates that testing is complete, and the result appears in the display. If you have not already flagged this quality control result, you may do so now. The value displayed here is an example.
6.3 Quality control test procedure 씰 If you performed the test with Accu-Chek Active Control 1, refer to row 1 for the permitted range. 씰 If you performed the test with Accu-Chek Active Control 2, refer to row 2 for the permitted range. ACCU-CHEK® Active Control mg/dL 1 36 – 66 2 131 – 177 If the result is within the stated range, you need to still carry out a visual plausibility test. It is important that this check be performed within 30–60 seconds after control solution was applied.
6.3 Quality control test procedure 씰 Turn over the test strip to reveal the round control window on the back. 씰 On the label of the test strip container is a colour scale with blood glucose values printed alongside. Select the blood glucose value that best approximates the reading you obtained. 씰 Compare the colour of the control window with the colour you selected on the label. The colours must be a fairly close match. If there is a great disparity, repeat the test.
6.4 Performing a full display check longer than 3 seconds The standard power-on display test checks the most important display elements. To verify that all of the elements are functioning correctly, you can carry out a full test. 씰 Press and hold down the M and S buttons together for longer than 3 seconds. 씰 All the elements of the liquid crystal display (LCD) are shown at once. Either “mg/dL” or “mmol/L” will be visible depending on the country-specific setting.
7 Keeping your monitor in tip-top condition Your Accu-Chek Active monitor has no moving parts and so will not suffer any mechanical wear and tear. As with any precision instrument, however, you will need to look after it carefully to keep it at its best. 7.1 Cleaning your Accu-Chek Active monitor Accu-Chek Active utilises an optical measuring method that relies heavily on all of its components being clean.
7.1 Cleaning your Accu-Chek Active monitor Clean the measurement optics carefully with nothing other than cold water, soft lint-free cloths and cotton swabs. For disinfection you may use 70 % alcohol. Any other cleaning agents may damage the monitor or impair its measuring function. 씰 Slide off the test strip guide towards you (see illustration). 씰 After removal of the test strip guide from the monitor, clean it with cold water. 씰 Afterwards you may wipe the test strip guide with 70% alcohol.
7.1 Cleaning your Accu-Chek Active monitor 씰 Wipe the measurement optics components with a soft lint- free cloth and/or a cotton swab. The cloth/cotton swab may be slightly moistened with water or 70% alcohol. Make sure that no liquid enters the monitor itself. Avoid scratching the measurement optics, as this will impair the measuring function. 씰 When all of the components are thoroughly dry, you may slide the test strip guide back onto the monitor. Ensure it clicks back into place.
7.1 Cleaning your Accu-Chek Active monitor Cleaning and disinfection in the professional area Be sure to observe all relevant Health and Safety regulations. For cleaning use only cold water and cotton swabs. 6 % formaldehyde solution may be used for disinfection. Please take into account the time it takes for the agents used to act. Any other cleaning agents may damage the monitor or impair its measuring function. A potential infection risk exists.
7.2 Battery life and battery replacement A new battery provides enough power for about 1,000 tests. When you first see the battery symbol displayed, there is still enough power for about 50 tests, though we recommend that in this event you replace the battery as soon as possible. By then the battery will have been considerably drained, and changeable conditions (cold surroundings) can affect its performance even further.
7.2 Battery life and battery replacement 씰 Slide the battery compartment back into the monitor and press gently inwards until it clicks home. 7.3 Measurement and storage conditions When testing blood glucose, pay close attention to the following points. They affect the dependability of your Accu-Chek Active monitor and the accuracy of your results. 씰 After replacing the battery, check the performance of the monitor as described in Chapter 6.
7.3 Measurement and storage conditions Do not use results obtained at the limits of the permitted temperature range as a basis for therapeutic decisions. Atmospheric humidity Never try to speed up warming or cooling of your Accu-Chek Active monitor artificially (e.g. by placing it on a radiator or in a refrigerator). Doing so can damage your monitor and cause it to give incorrect results.
8 Summary: All the display messages on your Accu-Chek Active monitor 8.2 Messages and symbols 8.1 About the display messages When you use your Accu-Chek Active monitor, various letters and symbols will appear on the display, at times also error messages. Some of these display messages, those occurring in routine use, have already been explained. We now present a complete list of display messages, what they mean, and what action you should take in case you see an error message.
8.2 Messages and symbols This message appears ... ● ● in Test Mode when the result is displayed (Test and Memory) and means: ● The ambient temperature is/was close to the limit at the time of testing. Do not use the result as a basis for therapeutic decisions. This message appears ... ● during testing ● The test is in progress. ● during testing, when enabled attached to results in memory ● This is/was a quality control test. ● in Memory Mode ● The figure is the 7/14 day average.
8.2 This message appears ... ● 8.3 Troubleshooting Messages and symbols in Test Mode and means: ● (flashing) in Test Mode ● ● in Test Mode ● ● (flashing) in Test Mode ● ● 116 015134 AccuChek Active_Eng.indd 116-117 The light is too bright, testing is not possible. Please move to another location. Apply a drop of blood now. This error message means ... The test strip has been inserted incorrectly.
8.3 Troubleshooting Measurement optics error. Clean the measurement optics (Chapter 7.1). Check that the test strip is lying flat and straight in the test strip guide. If the message persists, the monitor has a fault. Please call your customer support and service centre. Although rarely occuring - you applied too little blood to the strip. Repeat the test with a new Accu-Chek Active test strip. The measurement optics are dirty or you have inserted a test strip that has already been used or is expired.
8.4 Possible sources of error If you receive frequent error messages or if you often obtain implausible results, please check the following: 씰 Are the test strips and/or control solution still within the stated 씰 With your Accu-Chek Active monitor, are you using only 씰 Have you cleaned your Accu-Chek Active monitor (especially Accu-Chek Active test strips? expiry date? the test strip guide and measurement optics) as instructed in Chapter 7.
9 Appendix 9.1 Technical data Monitor type Accu-Chek Active Automatic power-off 1–2 minutes after last button press Catalogue number/ serial number See type plate on the back of the monitor Power supply 1 lithium battery type CR2032 or type DL2032 Test principle Determination of glucose in fresh capillary blood by reflectance photometry. When using different specimen material please refer to the package insert that came with the Accu-Chek Active test strips. Battery life approx.
9.2 System components With your Accu-Chek Active monitor use only Accu-Chek Active test strips. These are available from your pharmacist. For healthcare professionals we recommend the Accu-Chek Softclix Pro lancing device and special lancets Accu-Chek Softclix Pro Lancet. For quality control testing of your Accu-Chek Active monitor use only Accu-Chek Active Control solutions (see Chapter 6).
9.3 Guarantee 9.4 Local customer support and service The statutory guarantee provisions governing the sale of consumer goods in the country of purchase shall prevail. Customer support and service Customer support: If you have any questions on how to use your monitor, if any of the results you obtain seem implausible, or if you suspect the monitor has a fault, please contact your customer support centre. See the next Chapter for addresses.
9.5 Addresses Argentina Productos Roche S.A.Q. e I. DIAGNOSTICS DIVISION Rawson 3150, B1610BAL Ricardo Rojas Partido de Tigre (Buenos Aires) phone: +54 11 5129-8000/8457 fax: +54 11 5129-8105 Chile Productos Roche Ltda. Avda. Quilín 3750 Macul, Santiago Tel: 56 (2) 441 3200 Fax: 56 2 4626687 Línea gratuita: 800 471 800 Australia Roche Diagnostics Australia Pty Ltd. 31 Victoria Avenue, Castle Hill, NSW 2154, Australia Extracare Enquiry line: 1800 251816 www.rochediagnostics.com.
9.5 Addresses Malaysia Roche Diagnostics (M) Sdn Bhd Infinite Centre 1, 1st Floor, Lot 1, Jalan 13/6 46200 Petaling Jaya Selangor Darul Ehsan Toll-Free Line: 1-800-88-1313 Mexico Productos Roche S.A. de C.V. (Roche Diagnostics México) Av. Paseo de la Reforma 2620 Piso 8, Col. Lomas Altas 11950 México D.F. Teléfono: +52 (55) 50 81 58 00 Sin Costo 01800 90806 00 Fax: +52 (55) 50 81 58 58 Pakistan Roche Pakistan Ltd. 37 - C, Block 6 P.E.C.H.S. P.O.BOX 20021 Karachi - 75400 Tel.
10. Alphabetical index Ambient temperature Atmospheric humidity Average value Battery Battery capacity Blood application Blood collection Causes of error Cleaning the monitor Coding Coding chip Colour comparison Colour scale Control solution Control value Control window Customer support Data downloading Display Display test, full Disposal Drop symbol Electromagnetic interference 132 015134 AccuChek Active_Eng.indd 132-133 114, 119 111, 123 10, 67ff., 93, 115 5, 18, 81, 106ff. 11, 114, 123 10, 43ff.
10. Alphabetical index Saving results Service Set Mode Setting the date Setting the time Setting the year Shut-off time Shut-off, automatic Special testing situations Storage conditions Sun symbol Symbols System components Technical data Temperature range Test strip guide Testing by healthcare professionals Thermometer symbol Turning on Underdosing Units Warnings 60f 51, 87, 25ff. 17, 23ff 25ff. 25ff. 29 11, 39, 85, 123 47, 73, 93, 97, 123 55, 59 109ff., 120 110, 116 11, 112, 113ff. 124ff. 122ff.
Warranty Card Name Street City/post code Telephone No. Date Signature 015134 AccuChek_Garantie_Eng.indd 1 12.08.
Accu-Chek Active Warranty Card Please let this card be completed by the dealer and preserve it carefully. If this card was not filled out at the time of purchase please submit the original receipt with your claim. Serial-No. The statutory guarantee provisions governing the sale of consumer goods in the country of purchase shall prevail. Date of purchase Stamp and signature of dealer 015134 AccuChek_Garantie_Eng.indd 2 12.08.