Product Manual
Introduction
19
Battery pack
The battery pack begins charging as soon as the power adapter is connected or the meter is
placed on an active Handheld Base Unit (i.e., one connected to a power supply).
When storing or disposing of the battery pack, use the manufacturer's original packaging.
WARNING
Possible hazards posed by the battery pack
Damaged or swollen battery packs can overheat, catch fire, or leak.
Immediately cease use of CoaguChek XS Pro meters with damaged or
swollen battery packs and under no circumstances recharge them (do
not place in the base unit).
Overheating can cause the battery pack to catch fire or explode.
■ Never throw the battery pack or the meter onto a fire. Do not disman-
tle, compress, or pierce the battery pack as this could cause an inter-
nal short circuit that leads to overheating.
■ Do not place either the battery pack or the CoaguChek XS Pro meter
on or in heating appliances, such as a microwave, conventional oven,
or radiator.
■ Avoid prolonged exposure to direct sunlight, e.g., when the meter is
docked in the base unit. Keep this in mind when positioning the base
unit.
Battery fluid or materials leaking from damaged battery packs can irritate
your skin or cause burns due to high temperatures.
■ Avoid contact with leaking battery fluid. In the event of accidental
contact with the skin, rinse with water. If you get battery fluid in your
eye(s), you should also seek medical attention.
Handle and dispose of battery packs with care.
Extreme temperatures reduce the charging capacity and usage period of the meter and the
battery pack.
NOTICE
Use only the specially designed battery pack provided by Roche
Diagnostics. Using any other type of battery may damage the system.
Save or download data from the meter prior to replacing the battery pack to prevent loss of
data (see Chapter 8).