Product Manual
6 Operator’s Manual V03
3.3 Menu Functions
Self Check: During Self Check at power-on, the analyzer automatically checks that the program chip, tray transport
mechanism, printer connection and optical system are all operating properly. The tray type is checked
(see Sections 4 and 7.1) to ensure that it correctly matches the test strip type selected in the menu.
Calibration: For requesting calibration with Chemstrip Calibration Strip (see Section 5).
Linefeed: Causes paper to advance. The linefeed is stopped by pressing the left function key (“Stop”) (see Section 4).
New Series: For starting a new series of measurements at sequence number 1. It is also possible to have an Automatic
Patient ID (see Section 6.1).
Memory: The analyzer can store up to 100 results together with date and time of measurement, sequence number and
patient ID (if entered). Memory is automatically cleared every time the date changes.
After memory has been erased, “NO RESULTS STORED” is displayed when “Memory” is pressed. Pressing the START
button resumes Ready-to-Measure status. “MEMORY FULL” and the options “Print/Send/Clear” are displayed when the
memory is full. Memory must be cleared before the analyzer can resume Ready-to-Measure status.
Print Results: For generating a printout of stored results. The options are as follows:
• All: All results in memory (i.e. for current date)
• Last Series: The most recent series of readings
• Last one: The most recent reading
Printing can be repeated as often as desired. Printing can be interrupted by pressing the left key (“Stop”), for example to
allow a new roll of printer paper to be inserted (“Linefeed”), and subsequently resumed (“Continue”). The analyzer resumes
Ready-to-Measure status when the “Home” key is pressed or when printing has finished.
Send Results: For sending stored results to the serial interface. Options are the same as for “Print Results“. Results can be
sent as often as required (See Sections 6.7 and 9.1).
Clear Memory: Erases results from memory.
Mode: Choice of Print and Interface modes.
Printer: Printer options are:
• On: The printer is switched on. Each result is printed once.
• 2 Copies: Each result is printed twice. Note: When the printout is a repeat printout (activated by the Print Results
function), each result is printed only once.
• Off: For switching off the printer when the printout of results is only required at the end of a series of readings
(activated by the “Print Results” function), or when results are only to be sent via the interface to a personal or host
computer.
Interface: Choice of unidirectional, bidirectional or ASTM data transfer. For further details, see Sections 6.7 and 9.
Logdata: Choice of sending the log files to the host PC or printing out the last 10 logged data. (Only by supervisors in
Authentication mode)
Strip Settings: For selecting test strip type and units.
Strip Type: Choice of test strip type:
• Chemstrip 10 MD* Test Strip (CHEM 10)
• Chemstrip 5 OB Test Strip (CHEM 5)
• Chemstrip 7 Test Strip (CHEM 7)
The Urisys
1100
®
Urine Analyzer leaves the factory configured for Chemstrip 10 MD Test Strips. If you intend to use
Chemstrip 5 OB or Chemstrip 7 Test Strips, you will need to use the appropriate test strip tray (see Sections 4.3 and 11).