E14 INTERFACE ERROR ! Cause: Fault in data transfer to PC or host in bidirectional or ASTM mode.
Action: Check the data cable. Verify that the PC or host is ready to receive data.
Use the “Send” function to transfer data or press “Home” to resume Ready-to-
Measure status.
E17 INVALID PASSWORD! Cause: The entered password doesn’t match.
Action: Enter an correct password.
E18 INVALID OP.ID! Cause: The entered Operator ID is not valid.
Action: Enter a valid Operator ID.
E19 LIST DOWNLOAD FAILED Cause: The new Operator ID list download failed.
Action: No action. After 2 seconds next state starts with old list if there was.
E20 NO VALID LIST! Cause: There is not a valid list at all in device.
Action: Try to download an Operator ID list from the host or continue without
authenticated operator.
CHECK MEASUREMENT Analyzer prints out software and chip version number and 3-digit numbers
without naming the parameters.
Cause: Service function is activated.
Action: Press the “Back” function key to return to the main menu.
VALUES OBTAINED ARE Cause: Test strip incorrectly positioned or uncharacteristic test pad colors.
IMPLAUSIBLE WHEN The wrong test strip, may have been used. Electromagnetic interference from
COMPARED WITH THOSE FROM other devices (see Section 4).
VISUAL EVALUATION Action: Repeat the measurement with a new test strip. Follow the directions
carefully and ensure the test strip is correctly inserted. Repeat calibration if
necessary. Remove external sources of interference, if there are any.
NO PRINTOUT Cause: “Printer: Off” has been selected, or the printer/software is defective, or
the printer is out of paper.
Action: Insert paper if needed. Choose “Printer: On” to re-activate the printer.
Request a patient report via the “Print” function. If this fails, activate the
“Linefeed” function.
If there is still no response, contact Roche Diagnostics Technical Service at
1-800-428-4674.
THE ANALYZER WILL NOT READ Action: If an AT/PC keyboard is connected, press the Escape key, or switch the
EVEN THOUGH THE SEQUENCE Urisys
1100
®
Urine Analyzer off and back on again.
NUMBER IS DISPLAYED
9. Connecting to Other Devices
9.1 Serial Interface
At the rear, the Urisys
1100
®
Urine Analyzer has a serial interface through which it can be connected to a PC or central host
computer. This is not an RS 232 type interface.
Roche Diagnostics has a suitable standard data cable available for sale (see Section 11).
The interface can be used for unidirectional, bidirectional or ASTM communication, selected via the menu.