Product Manual

16 Operator’s Manual V03
3. Press the START button (see Figure 12). An acknowledging beep sounds. The
tray advances slightly, the retaining bar closes, and the gray reference pad on
the tray is read (see Figure 13).
4. If the test strip is not correctly located in the middle of the tray, move it gently
to the side until it is properly aligned (see Figure 14). Be careful not to move
the tray.
5. Seventy seconds after the START button is pressed, all test strip pads are
read. The results are printed and the next sample number appears on the
display.
6. The test strip tray returns to the starting position and the retaining bar opens.
Remove and dispose of the test strip. Wipe any urine residue from the tray
with a lint-free cloth (see Figure 15).
6.4 Patient Report
The patient report is printed out together with the sequence number, device ID,
operator ID, date and time. The patient’s name will also appear on the print out if
entered prior to measurement (see Section 6.6). Test results which diverge from
negative, normal, or trace values are flagged with an asterisk before the
parameter concerned. For details regarding various print settings and how to
modify them, see Section 3.3.
Tear off the printout, by pulling it horizontally over the edge.