4.2 Installation and Power-on Procedure
Set-Up
(1) Place the Urisys
1100
®
Urine Analyzer on a stable, level surface. To ensure accurate readings, please take the following
environmental precautions:
!
Do not set the analyzer in close proximity to devices that create high-frequency fields, as they may interfere and
produce false results. Such devices include; walkie-talkies, mobile telephones, microwave ovens and diathermic
equipment.
Do not expose the analyzer to direct sunlight or strong artificial light.
If the analyzer has been exposed to significant changes in temperature and/or humidity, allow it to sit at room
temperature for at least four hours before operating.
(2) Connect the Power cord and AC adaptor. Then connect the AC adaptor to the power socket at the rear of the
instrument. Plug the power cord into a readily accessible AC power outlet.
Insertion of Test Strip Tray
(3) Select the appropriate test strip tray. Use Type C to read Chemstrip 10 MD*
Urine Test Strips. Use Type N to read either Chemstrip 5 OB or Chemstrip 7
Urine Test Strips. The type of tray is indicated on the underside of the tray.
See Section 4.3 for details on using the Urisys 1100
®
Urine Analyzer with
Chemstrip 5 OB or Chemstrip 7 Urine Test Strips.
Hold the test strip tray with the gray reference pad facing up and towards
the analyzer (see Figure 1). Slide the test strip tray into the slot below the
function keys until the retaining bar closes (see Figure 2).
!
Be careful not to touch the gray reference pad. Contamination of the reference pad may impair the quality
of the results obtained.
Figure 1
Figure 2
* Hospitals may use Chemstrip 10 UA Urine Test Strips (Cat. No. 418003).