Product Manual

7
Operator’s Manual V03
Units: Options are:
Conventional units (mg/dL)
SI units (mmol/L)
Arbitrary units (+, ++, +++, ++++)
The operator selects the units in which the results are to be stored, printed and/or transferred to a computer. After a new
unit setting has been selected, the repeat printout (activated by “Print Results”) and all following printouts and/or data
transfer (activated by “Send Results”) will be in the newly chosen units.
Language: This enables the display language to be set. The “Other” option allows the operator to choose between English,
German, Italian, Spanish and French.
Device ID: Displays the 5 digits device ID, which is part of the factory settings and cannot be changed.
The device ID will also be sent to the host.
Operator ID: Choice between Normal and Authentication modes for the operator identification. If activated the input of an
operator ID will be required by the start of the instrument.
The Operator ID will appear in the result printout and will also be sent to the host PC. Authentication mode offers a lock-
out function and requires the ASTM protocol. (refer Section 4.2 and 6.6 for details.)
DATE/TIME: For setting the date and time.
The factory default is the date in Month-Day-Year order and the time in hours (12-hour clock) and minutes. If required, the
time can be displayed in the “24-hour clock” mode. Pressing the “Sequence” key allows the date format to be changed to
Day-Month-Year or Year-Month-Day. “Set” causes the date and time to be displayed and set. Pressing the left function key
(<<<) moves the blinking cursor to the left. The time or date unit highlighted by the cursor can then be increased or
decreased by pressing the + / — keys.
Pressing the START button confirms the setting, closes the submenu, and resumes Ready-to-Measure
status.
3.4 Default Settings
The Urisys 1100
®
Urine Analyzer comes equipped with the following default settings. These can be changed using the
function keys to page through the menus and the START button to activate the new settings selection. (See Sections 3.2
and 3.3.)
Parameter Default Setting Optional Setting
Print results Printer On, (one copy) Printer On (two copies)
Printer Off
Interface mode Uni-directional Bi-directional
Operator ID Off On / Authentication Mode (ASTM)
Language English German, Italian, Spanish, French
Date/time format Date: Month-Day-Year
Time: 12-hour clock
Date: Day-Month-Year / Year-Month-Day
Time: 24-hour clock
Reporting Unit Conventional units SI units or Arbitrary Units
Test Strip Type Chemstrip 10 Chemstrip 5 OB
Chemstrip 7
Follow the flowchart in section 3.2 to determine the menu pathway to a particular setting. Use the function keys and the
START button to access and change the setting.
Please Contact Roche Technical Service at 1-800-428-4674 if you need assistance.
* Hospitals may use Chemstrip 10 UA Urine Test Strips (Cat. No. 418003).