Cleaning and Disinfecting the Meter and the Lancing Device
CoaguChek INRange User’s Manual
3 Hold the meter upright with the test strip
guide facing down.
• Clean the easily accessible white areas
with a moistened cotton swab/bud.
• Ensure the swab/bud is only damp, not
wet. Wipe away residual moisture and
uids.
NOTICE
Damage to the instrument
• Ensure that no liquid enters the meter. If
moisture enters the meter, it may cause
malfunction of the instrument.
• Do not insert any objects in the test strip
guide. Doing so may damage the electrical
contacts behind the test strip guide.