Users Manual Part 2

283Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Related L1 message
I-AC
u About level 1 messages and level 2 errors (209)
MS-0020: Test strip ejection error
Problem
The measuring unit failed to eject the test strip.
Solution
Eject the test strip using the eject button.
If the problems persists, remove the test strip manually.
Related L1 message
H-AG
u About level 1 messages and level 2 errors (209)
MS-0021: Measuring unit error
Problem
A measuring unit error occurred.
Solution
Restart the instrument and try again.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
MS-0022: SW update error
Problem
Software update was not completed successfully.
Solution
Trigger the software update process again via
cobas
®
infinity edge.
Related L1 message
I-AC
u About level 1 messages and level 2 errors (209)